FDA Adverse Event Malfunction Summary report: N

ID NOW COVID-19 ASSAY

MDR report key: 11847380 · Received May 19, 2021

Report

Report Number
1221359-2021-01529
Event Type
Malfunction
Date Received
May 19, 2021
Report Date
August 16, 2021
Manufacturer
ABBOTT DIAGNOSTICS SCARBOROUGH, INC
Product Code
QJR
UDI-DI
10811877011269
PMA / PMN Number
EUA2000074
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

THIS INVESTIGATION IS STILL IN PROGRESS. ONCE THE INVESTIGATION IS COMPLETE A SUPPLEMENTAL REPORT WILL BE PROVIDED. (B)(6). REFERENCE MFR. REPORT: 1221359-2021-01488.

Additional Manufacturer Narrative · 0

ADDITIONAL INFORMATION: TESTING WAS PERFORMED AT ABBOTT DIAGNOSTICS (B)(4), INC. ON RETAINED KIT LOT M141938 WITH INTERNAL POSITIVE QUALITY CONTROL SAMPLES AND NEGATIVE QUALITY CONTROL SWABS. ALL TEST RESULTS WERE VALID AND PERFORMED AS EXPECTED. ADDITIONALLY, THE MANUFACTURING RECORDS AND QUALITY CONTROL RELEASE TESTING WAS REVIEWED FOR TEST KIT PART NUMBER 190-000 / LOT: M141938, TEST BASE PART NUMBER 190-430 / LOT: M141938. THE LOT MET THE REQUIRED RELEASE SPECIFICATIONS. A REVIEW OF THE COMPLAINTS REPORTED AS FALSE POSITIVE PATIENT RESULTS (CONFIRMED AND UNCONFIRMED, CONFLICTING RESULTS) RELATED TO KIT LOT M141938 SHOWED THAT THE COMPLAINT RATE IS (B)(4). ABBOTT DIAGNOSTICS (B)(4) WAS UNABLE TO DETERMINE THE EXACT ROOT CAUSE OF THE REPORTED ISSUE HOWEVER A POSSIBLE ASSIGNABLE ROOT CAUSE IS SAMPLE INTERFERENCE OR CROSS CONTAMINATION.

Description of Event or Problem · 0

THE CUSTOMER REPORTED 3 FALSE POSITIVE RESULTS WITH THE ID NOW COVID-19 ASSAY (DIFFERENT LOT NUMBERS). THIS MFR. ADDRESSES LOT NUMBER 2 OF 2. NO ADDITIONAL INFORMATION, INCLUDING PATIENT TREATMENT AND OUTCOME HAVE BEEN PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
748401 ID NOW COVID-19 ASSAY MOLECULAR IVD FOR ID NOW COVID-19 QJR ABBOTT DIAGNOSTICS SCARBOROUGH, INC M141938 10811877011269

Patients

Seq Age Sex Outcome Treatment
1