FDA Adverse Event Malfunction Summary report: N

DCTM

MDR report key: 1184181 · Received September 26, 2008

Report

Report Number
1418964-2008-00001
Event Type
Malfunction
Date Received
September 26, 2008
Date of Event
August 27, 2008
Report Date
September 26, 2008
Manufacturer
DEL MEDICAL IMAGING CORP.
Product Code
IYB
PMA / PMN Number
NA
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

PRELIMINARY INDICATIONS ARE THAT THE MALFUNCTION MAY HAVE BEEN DUE TO POOR MAINTENANCE. THE UNIT IS OUT OF SERVICE AND IS BEING RETURNED FOR EVALUATION, AND ANY ADDITIONAL RESULTS OR INFORMATION WILL BE PROVIDED IN A FOLLOW-UP REPORT.

Description of Event or Problem · 1

COUNTER WEIGHT CABLE CAME LOOSE DUE TO WORN ANCHOR PIN IN COLUMN CAUSING COLUMNS TO DROP DOWN.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 DCTM X-RAY CEILING TUBE MOUNT IYB DEL MEDICAL IMAGING CORP. DCTM NA

Patients

Seq Age Sex Outcome Treatment
1