FDA Adverse Event Malfunction Summary report: N

CADD CASSETTE RESERVOIR

MDR report key: 11841773 · Received May 18, 2021

Report

Report Number
3012307300-2021-04517
Event Type
Malfunction
Date Received
May 18, 2021
Date of Event
April 19, 2021
Report Date
December 15, 2022
Manufacturer
ST PAUL
Product Code
FPA
UDI-DI
10610586027239
PMA / PMN Number
K040636
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
KS, US
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

OTHER, OTHER TEXT: THIS REMEDIATION MDR WAS GENERATED UNDER PROTOCOL B10010116, AS A RESULT OF WARNING LETTER (B)(4). A DEVICE HISTORY RECORD (DHR) REVIEW WAS CONDUCTED FOR ALL LOTS AND ALL INSPECTIONS WERE COMPLETED AND NO ISSUES WERE NOTED DURING MANUFACTURE. THREE CASSETTE UNITS WERE RECEIVED FOR EVALUATION. THEY WERE RECEIVED IN NEW CONDITION WITH ORIGINAL PACKAGE CLOSED. VISUAL AND FUNCTIONAL TESTING WERE PERFORMED. VISUAL INSPECTION FOUND THERE WAS NO DAMAGE, SCUFFS, PINCH MARKS, CRACKS, CRAZING, ETC. COULD CAUSE THE FAILURE MODE REPORTED. FUNCTIONAL REVEALED THAT THE SAMPLES WERE FULLY PRIMING AND CONNECTED WITHOUT DIFFICULTLY; THE PUMP RAN AND THE ALARM WAS NOT ACTIVATED. THE COMPLAINT WAS NOT CONFIRMED. THE ROOT CAUSE OF THE ISSUE COULD NOT BE DETERMINED. THIS ISSUE WILL BE MONITORED FOR ANY INCREASE IN OCCURRENCE. IF THE OCCURRENCE OF THIS ISSUE INCREASES SIGNIFICANTLY, ADDITIONAL INVESTIGATIVE ACTION WILL BE TAKEN.

Additional Manufacturer Narrative · 1

POTENTIAL LOT NUMBERS INVOLVED ARE 4092492, 4092493, 4060911, 4076292. ADDITIONAL CONTACT: (B)(6). PRODUCT COMPLAINT ANALYST.

Description of Event or Problem · 1

INFORMATION WAS RECEIVED INDICATING THAT A SMITHS MEDICAL CADD CASSETTE RESERVOIR CAUSED "NO DISPOSABLE" ALARMS. THERE WAS NO PATIENT INJURY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
739441 CADD CASSETTE RESERVOIR SET, ADMINISTRATION, INTRAVASCULAR FPA ST PAUL 21-7302-24 10610586027239

Patients

Seq Age Sex Outcome Treatment
1 Unknown