ELECSYS IL 6
Report
- Report Number
- 1823260-2021-01477
- Event Type
- Malfunction
- Date Received
- May 18, 2021
- Date of Event
- April 26, 2021
- Report Date
- June 11, 2021
- Manufacturer
- ROCHE DIAGNOSTICS
- Product Code
- QLC
- PMA / PMN Number
- EUA
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- RS
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
THE CUSTOMER'S CALIBRATION RESULTS WERE OK BUT LOWER THAN THE EXPECTED RANGES. THE CUSTOMER'S QC RESULTS FROM (B)(6)2021 WERE OK. THE INVESTIGATION REVIEWED THE SYSTEM'S ALARM TRACE AND NO CRITICAL ALARMS WERE DISCOVERED. TWO PATIENT SAMPLES WERE PROVIDED FOR AN INVESTIGATION. THE INVESTIGATION DISCOVERED THE PATIENT'S SAMPLE COLLECTED ON (B)(6)2021 WAS HEMOLYTIC. THE INVESTIGATION TESTED THE PROVIDED SAMPLES AND THE CUSTOMER'S RESULTS WERE REPRODUCED. NO INTERFERING FACTOR WAS FOUND IN EITHER OF THE PROVIDED SAMPLES. THE INVESTIGATION DID NOT IDENTIFY A PRODUCT PROBLEM. THE CAUSE OF THE EVENT COULD NOT BE DETERMINED.
THIS EVENT OCCURRED IN (B)(6). UNIQUE DEVICE IDENTIFIER (UDI) (B)(4). THE INVESTIGATION IS ONGOING.
THE INITIAL REPORTER RECEIVED QUESTIONABLE ELECSYS IL 6 RESULTS FOR ONE PATIENT TESTED ON A COBAS E411 DISK SERIAL NUMBER (B)(4). THE PATIENT'S INITIAL IL SIX (6) RESULT WAS NOT REPORTED OUTSIDE THE LABORATORY. THE CUSTOMER PERFORMED REPEAT TESTING WITH A VECTOR-BEST ELISA ANALYZER AND THE COBAS E411 DISK. ON (B)(4) 2021, THE PATIENT'S ELECSYS INITIAL IL SIX (6) RESULT WAS 1.50 PG/ML WITH A DATA FLAG. THE PATIENT'S IL 6 RESULT WITH THE SAME SAMPLE ON THE VECTOR-BEST ELISA ANALYZER WAS 300 PG/ML. ON (B)(6) 2021, THE PATIENT'S REPEAT ELECYS IL SIX (6) RESULT WITH THE SAME SAMPLE ON THE COBAS E411 DISK WAS 1.77 PG/ML. ON (B)(6) 2021, THE PATIENT HAD A NEW SAMPLE COLLECTED AND THE PATIENT'S ELECSYS IL SIX (6) RESULT ON A COBAS E411 DISK WAS 2.60 PG/ML.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 736294 | ELECSYS IL 6 | INTERLEUKIN 6 | QLC | ROCHE DIAGNOSTICS | IL-6 | 49651501 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |