FDA Adverse Event Malfunction Summary report: N

630G INSULIN PUMP MMT-1715KL 630G

MDR report key: 11840346 · Received May 18, 2021

Report

Report Number
2032227-2021-146412
Event Type
Malfunction
Date Received
May 18, 2021
Date of Event
April 14, 2021
Report Date
April 26, 2022
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OZO
UDI-DI
000000763000166519
Removal / Correction Number
Z-0955-2020
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
VA, US
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

THIS REPORT IS PART OF A RETROSPECTIVE REVIEW AND REMEDIATION EFFORTS IN RESPONSE TO A WARNING LETTER. UPDATED H9: Z-0955-2020. MEDTRONIC, INC. (MEDTRONIC) IS SUBMITTING THIS REPORT TO COMPLY WITH 21 C.F.R. PART 803, THE MEDICAL DEVICE REPORTING REGULATION. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION IN THE TIME ALLOTTED AND HAS PROVIDED AS MUCH INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY "DEFECTS" OR HAS "MALFUNCTIONED". THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA, TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REPORTING PURSUANT TO PART 803. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE IT BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT.

Additional Manufacturer Narrative · 1

(B)(4). THE INSULIN PUMP WAS RECEIVED WITH A SCRATCHED CASE AND A CRACKED RETAINER. THE TEST RESERVOIR DOES LOCK PROPERLY INSIDE THE RESERVOIR TUBE. THE INSULIN PUMP PASSED THE DISPLACEMENT TEST AND SELF TEST. HOWEVER, HARDWARE LOW LEVEL FAILURES ALARM CONFIRMED IN THE PUMP HISTORY DUE TO BROKEN TRACE ON KEYPAD ASSEMBLY.

Description of Event or Problem · 1

INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP HAD HARDWARE LOW LEVEL FAILURES ERROR ALARM. CUSTOMER WAS ABLE TO CLEAR THE ALARM AND ABLE TO COMPLETE REWIND. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WAS RETURNED FOR ANALYSIS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
735801 630G INSULIN PUMP MMT-1715KL 630G ARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND OZO MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-1715KL HG3K00L 000000763000166519

Patients

Seq Age Sex Outcome Treatment
1 74 YR Female