L-CATH PICC S/L 28GA (1.2F) X 25CM 1 LUMEN
Report
- Report Number
- 0001625425-2021-00917
- Event Type
- Malfunction
- Date Received
- May 17, 2021
- Date of Event
- January 22, 2021
- Report Date
- April 20, 2021
- Manufacturer
- ARGON MEDICAL DEVICES
- Product Code
- FOZ
- UDI-DI
- 00886333209910
- PMA / PMN Number
- K091670
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
ONE DEVICE WAS RETURNED FOR REVIEW. A FUNCTIONAL TEST FOUND LEAKAGE THROUGH A CRACK IN THE HUB, CONFIRMING THE COMPLAINT. CAPA C-2020-016 HAS BEEN INITIATED TO IDENTIFY ROOT CAUSE AND CORRECTIVE ACTION. THE MOLDING PARAMETERS OF THESE HUBS WERE UPDATED TO MEET MANUFACTURERS SUGGESTED SETTINGS AND MOLD 501-05 CAVITY 4 HAD A NEW CORE PIN INSTALLED THAT CONTROLS ALL THE CRITICAL DIMENSIONS FOR THE INNER DIAMETER. THE CAPA IS IN THE EFFECTIVENESS CHECK STAGE.
ONE DEVICE WAS RETURNED FOR REVIEW. A FUNCTIONAL TEST FOUND LEAKAGE THROUGH A CRACK IN THE HUB, CONFIRMING THE COMPLAINT. CAPA C-2020-016 HAS BEEN INITIATED TO IDENTIFY ROOT CAUSE AND CORRECTIVE ACTION. THE MOLDING PARAMETERS OF THESE HUBS WERE UPDATED TO MEET MANUFACTURERS SUGGESTED SETTINGS AND MOLD 501-05 CAVITY 4 HAD A NEW CORE PIN INSTALLED THAT CONTROLS ALL THE CRITICAL DIMENSIONS FOR THE INNER DIAMETER. THE INVESTIGATION IS STILL IN PROGRESS.
CALLED TO PATIENTS BEDSIDE BECAUSE PICC DRESSING WAS WET. LINE FOUND TO BE LEAKING WHEN FLUSHED.
CALLED TO PATIENTS BEDSIDE BECAUSE PICC DRESSING WAS WET. LINE FOUND TO BE LEAKING WHEN FLUSHED.
CALLED TO PATIENTS BEDSIDE BECAUSE PICC DRESSING WAS WET. LINE FOUND TO BE LEAKING WHEN FLUSHED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 735524 | L-CATH PICC S/L 28GA (1.2F) X 25CM 1 LUMEN | L-CATH PICC | FOZ | ARGON MEDICAL DEVICES | 384516 | 11308071 | 00886333209910 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Required Intervention |