FDA Adverse Event Malfunction Summary report: N

L-CATH PICC S/L 28GA (1.2F) X 25CM 1 LUMEN

MDR report key: 11838528 · Received May 17, 2021

Report

Report Number
0001625425-2021-00917
Event Type
Malfunction
Date Received
May 17, 2021
Date of Event
January 22, 2021
Report Date
April 20, 2021
Manufacturer
ARGON MEDICAL DEVICES
Product Code
FOZ
UDI-DI
00886333209910
PMA / PMN Number
K091670
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

ONE DEVICE WAS RETURNED FOR REVIEW. A FUNCTIONAL TEST FOUND LEAKAGE THROUGH A CRACK IN THE HUB, CONFIRMING THE COMPLAINT. CAPA C-2020-016 HAS BEEN INITIATED TO IDENTIFY ROOT CAUSE AND CORRECTIVE ACTION. THE MOLDING PARAMETERS OF THESE HUBS WERE UPDATED TO MEET MANUFACTURERS SUGGESTED SETTINGS AND MOLD 501-05 CAVITY 4 HAD A NEW CORE PIN INSTALLED THAT CONTROLS ALL THE CRITICAL DIMENSIONS FOR THE INNER DIAMETER. THE CAPA IS IN THE EFFECTIVENESS CHECK STAGE.

Additional Manufacturer Narrative · 0

ONE DEVICE WAS RETURNED FOR REVIEW. A FUNCTIONAL TEST FOUND LEAKAGE THROUGH A CRACK IN THE HUB, CONFIRMING THE COMPLAINT. CAPA C-2020-016 HAS BEEN INITIATED TO IDENTIFY ROOT CAUSE AND CORRECTIVE ACTION. THE MOLDING PARAMETERS OF THESE HUBS WERE UPDATED TO MEET MANUFACTURERS SUGGESTED SETTINGS AND MOLD 501-05 CAVITY 4 HAD A NEW CORE PIN INSTALLED THAT CONTROLS ALL THE CRITICAL DIMENSIONS FOR THE INNER DIAMETER. THE INVESTIGATION IS STILL IN PROGRESS.

Description of Event or Problem · 0

CALLED TO PATIENTS BEDSIDE BECAUSE PICC DRESSING WAS WET. LINE FOUND TO BE LEAKING WHEN FLUSHED.

Description of Event or Problem · 0

CALLED TO PATIENTS BEDSIDE BECAUSE PICC DRESSING WAS WET. LINE FOUND TO BE LEAKING WHEN FLUSHED.

Description of Event or Problem · 1

CALLED TO PATIENTS BEDSIDE BECAUSE PICC DRESSING WAS WET. LINE FOUND TO BE LEAKING WHEN FLUSHED.