BD EMERALD SYRINGE
Report
- Report Number
- 3002682307-2021-00194
- Event Type
- Malfunction
- Date Received
- May 17, 2021
- Date of Event
- April 19, 2021
- Report Date
- June 14, 2021
- Manufacturer
- BECTON DICKINSON, S.A.
- Product Code
- FMF
- PMA / PMN Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FI
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
H.6. INVESTIGATION: A DEVICE HISTORY RECORD REVIEW WAS PERFORMED FOR PROVIDED LOT NUMBER 2010206 AND THE REVIEW DID NOT REVEAL ANY DETECTED QUALITY ISSUES DURING THE PRODUCTION PROCESS THAT COULD HAVE CONTRIBUTED TO THIS REPORTED INCIDENT. AS SAMPLES WERE UNAVAILABLE FOR RETURN, TWENTY RETAINED SAMPLES OF THE SAME LOT NUMBER WERE OBTAINED FROM THE MANUFACTURING FACILITY FOR EVALUATION. THE RETAINED SAMPLES WERE EXAMINED AND NO SIGNS OF DEFECT WERE OBSERVED. ROOT CAUSE COULD NOT BE DETERMINED. H3 OTHER TEXT : SEE H.10.
IT WAS REPORTED THAT BD EMERALD¿ SYRINGE LEAKED FROM THE SPINAL NEEDLE ATTACHMENT DURING USE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "THE QUALITY OF EMERALD 2 ML SYRINGE HAS DECLINED. WHEN INJECTING FLUID WITH A SPINAL NEEDLE TO THE PATIENT, THE SYRINGE LEAKS FROM SYRINGE-NEEDLE ATTACHMENT POINT."
DATE OF EVENT: UNKNOWN. A DEVICE EVALUATION IS ANTICIPATED BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.
IT WAS REPORTED THAT BD EMERALD¿ SYRINGE LEAKED FROM THE SPINAL NEEDLE ATTACHMENT DURING USE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "THE QUALITY OF EMERALD 2 ML SYRINGE HAS DECLINED. WHEN INJECTING FLUID WITH A SPINAL NEEDLE TO THE PATIENT, THE SYRINGE LEAKS FROM SYRINGE-NEEDLE ATTACHMENT POINT."
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 734987 | BD EMERALD SYRINGE | PISTON SYRINGE | FMF | BECTON DICKINSON, S.A. | 2010206 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |