FDA Adverse Event Malfunction Summary report: N

THERMAGE CPT SYSTEM

MDR report key: 11838131 · Received May 17, 2021

Report

Report Number
3011423170-2021-00052
Event Type
Malfunction
Date Received
May 17, 2021
Date of Event
April 20, 2021
Report Date
April 20, 2021
Manufacturer
SOLTA MEDICAL, INC
Product Code
GEI
PMA / PMN Number
K132431
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NV, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

EVALUATION OF THE TREATMENT TIP CONFIRMED DAMAGE ALONG THE RADIOFREQUENCY TRACE OF THE TIP. THE INVESTIGATION FOUND DAMAGE TO THE RADIOFREQUENCY TRACE ARE CAUSED BY STRESS CONCENTRATIONS ON THE FLEX ASSEMBLY AT THE ADHESIVE EDGE THAT DAMAGE THE RADIOFREQUENCY TRACE, CAUSING ARCING AND SUBSEQUENT BURN-THROUGH OF THE FLEX CIRCUIT MEMBRANE. DEFECTS ON THE TIP MEMBRANE CAN CAUSE THE RADIOFREQUENCY ENERGY, DELIVERED BY THE SYSTEM, TO FOCUS IN A SMALL AREA OF THE MEMBRANE, RATHER THAN TO BE UNIFORMLY DISTRIBUTED OVER THE ENTIRE MEMBRANE AREA AND CAN POTENTIALLY CAUSE PATIENT BURNS. THIS CAN ALSO CAUSE SPARKING FROM THE TIP DURING TREATMENT CONFIRMING CUSTOMERS REPORT OF SPARKING DURING TREATMENT. A REVIEW OF THE MANUFACTURING RECORDS SHOWED ALL REQUIREMENTS WERE MET. FINAL TEST VERIFICATION SPECIFICATIONS ARE ACCEPTABLE. NO NON-CONFORMITIES OR ANOMALIES FOUND RELATED TO THIS COMPLAINT WHEN REVIEWING THE DEVICE HISTORY RECORD FOR SERIAL/LOT NUMBER (B)(6). A CAPA WAS OPENED FOR FURTHER INVESTIGATION.

Additional Manufacturer Narrative · 1

THE PRODUCT WAS RETURNED AND EVALUATED. A BURNT TRACE ON THE TIP MEMBRANE WAS CONFIRMED. THE TIP PASSED FLOW AND FAILED LEAK TESTS. THE TIP FAILED VISUAL INSPECTION FOR BURNT TRACE ON TIP SURFACE. DIELECTRIC BREAKDOWN WAS OBSERVED. THE TIP PASSED THERMISTOR TEST. FUNCTIONAL TESTING WAS NOT PERFORMED DUE TO DAMAGE ON THE TIP SURFACE. THE INVESTIGATION IS ONGOING.

Description of Event or Problem · 1

THE USER FACILITY REPORTED THAT THERE WAS A SPARK FROM THE TIP MEMBRANE. THIS WAS WITNESSED DURING THE TREATMENT PULSE. THE PULSES WERE PREVIOUSLY COMPLETED. THE USER ALSO STATED THAT SHE DIDN'T SEE ANY ABNORMAL PHYSICAL CONDITIONS ON THE SKIN WHERE SPARK OCCURRED AT END OF TREATMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
728832 THERMAGE CPT SYSTEM ELECTROSURGICAL,CUTTING & COAGULATION & ACC. GEI SOLTA MEDICAL, INC TTNS3.00E4-1200 222

Patients

Seq Age Sex Outcome Treatment
1