THERMAGE CPT SYSTEM
Report
- Report Number
- 3011423170-2021-00052
- Event Type
- Malfunction
- Date Received
- May 17, 2021
- Date of Event
- April 20, 2021
- Report Date
- April 20, 2021
- Manufacturer
- SOLTA MEDICAL, INC
- Product Code
- GEI
- PMA / PMN Number
- K132431
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NV, US
- Reporter Occupation
- OTHER
Narratives
EVALUATION OF THE TREATMENT TIP CONFIRMED DAMAGE ALONG THE RADIOFREQUENCY TRACE OF THE TIP. THE INVESTIGATION FOUND DAMAGE TO THE RADIOFREQUENCY TRACE ARE CAUSED BY STRESS CONCENTRATIONS ON THE FLEX ASSEMBLY AT THE ADHESIVE EDGE THAT DAMAGE THE RADIOFREQUENCY TRACE, CAUSING ARCING AND SUBSEQUENT BURN-THROUGH OF THE FLEX CIRCUIT MEMBRANE. DEFECTS ON THE TIP MEMBRANE CAN CAUSE THE RADIOFREQUENCY ENERGY, DELIVERED BY THE SYSTEM, TO FOCUS IN A SMALL AREA OF THE MEMBRANE, RATHER THAN TO BE UNIFORMLY DISTRIBUTED OVER THE ENTIRE MEMBRANE AREA AND CAN POTENTIALLY CAUSE PATIENT BURNS. THIS CAN ALSO CAUSE SPARKING FROM THE TIP DURING TREATMENT CONFIRMING CUSTOMERS REPORT OF SPARKING DURING TREATMENT. A REVIEW OF THE MANUFACTURING RECORDS SHOWED ALL REQUIREMENTS WERE MET. FINAL TEST VERIFICATION SPECIFICATIONS ARE ACCEPTABLE. NO NON-CONFORMITIES OR ANOMALIES FOUND RELATED TO THIS COMPLAINT WHEN REVIEWING THE DEVICE HISTORY RECORD FOR SERIAL/LOT NUMBER (B)(6). A CAPA WAS OPENED FOR FURTHER INVESTIGATION.
THE PRODUCT WAS RETURNED AND EVALUATED. A BURNT TRACE ON THE TIP MEMBRANE WAS CONFIRMED. THE TIP PASSED FLOW AND FAILED LEAK TESTS. THE TIP FAILED VISUAL INSPECTION FOR BURNT TRACE ON TIP SURFACE. DIELECTRIC BREAKDOWN WAS OBSERVED. THE TIP PASSED THERMISTOR TEST. FUNCTIONAL TESTING WAS NOT PERFORMED DUE TO DAMAGE ON THE TIP SURFACE. THE INVESTIGATION IS ONGOING.
THE USER FACILITY REPORTED THAT THERE WAS A SPARK FROM THE TIP MEMBRANE. THIS WAS WITNESSED DURING THE TREATMENT PULSE. THE PULSES WERE PREVIOUSLY COMPLETED. THE USER ALSO STATED THAT SHE DIDN'T SEE ANY ABNORMAL PHYSICAL CONDITIONS ON THE SKIN WHERE SPARK OCCURRED AT END OF TREATMENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 728832 | THERMAGE CPT SYSTEM | ELECTROSURGICAL,CUTTING & COAGULATION & ACC. | GEI | SOLTA MEDICAL, INC | TTNS3.00E4-1200 | 222 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |