FDA Adverse Event Malfunction Summary report: N

COLLEAGUE VOLUMETRIC INFUSION PUMP

MDR report key: 1183685 · Received October 4, 2008

Report

Report Number
6000001-2007-01378
Event Type
Malfunction
Date Received
October 4, 2008
Date of Event
December 14, 2006
Report Date
December 14, 2006
Manufacturer
BAXTER HEALTHCARE PTE. LTD.
Product Code
FRN
Removal / Correction Number
CONTINUED ON H10
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UT, US
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 1

THIS REPORT IS BEING RESUBMITTED IN ACCORDANCE WITH INSTRUCTIONS FROM FDA TO ADDRESS A MANUFACTURER REPORT SEQUENCE NUMBER ISSUE THAT OCCURRED WHEN THIS MDR OR SUPPLEMENT WAS ORIGINALLY SUBMITTED TO THE FDA ON JAN 12 2007. EVALUATION SUMMARY:THE REPORTED CONDITION OF DAMAGED BATTERIES WAS CONFIRMED BY THE BAXTER REPAIR TECHNICIAN. INSPECTION OF THE DEVICE REVEALED POTENTIALLY DAMAGED BATTERIES, WHICH WERE REPLACED. THE DEVICE WAS TESTED BY THE REPAIR TECHNICIAN AND RETURNED TO SERVICE. REVIEW OF THE COMPLAINT HISTORY REVEALS SIMILAR REPORTS HAVE BEEN RECEIVED FOR THIS PRODUCT FOR THE REPORTED ISSUE. THIS ISSUE IS BEING INVESTIGATED UNDER CAPA.

Description of Event or Problem · 1

THE DEVICE WAS RETURNED FOR SERVICE. DURING SERVICE TESTING, THE BAXTER TECHNICIAN REPORTED AN INFUSION PUMP WITH DAMAGED BATTERIES. PER THE SERVICE SHOP, THE HOSPITAL REPRESENTATIVE STATED THAT THERE HAVE BEEN NO REPORTS OF ANY PATIENT INJURY OR MEDICAL INTERVENTION. NO ADDITIONAL INFORMATION IS KNOWN.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 COLLEAGUE VOLUMETRIC INFUSION PUMP INFUSION PUMP FRN BAXTER HEALTHCARE PTE. LTD. NA NA

Patients

Seq Age Sex Outcome Treatment
1