COLLEAGUE CX VOLUMETRIC INFUSION PUMP
Report
- Report Number
- 6000001-2007-00776
- Event Type
- Malfunction
- Date Received
- October 4, 2008
- Date of Event
- December 14, 2006
- Report Date
- December 14, 2006
- Manufacturer
- BAXTER HEALTHCARE PTE. LTD.
- Product Code
- FRN
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
THIS REPORT IS BEING RESUBMITTED IN ACCORDANCE WITH INSTRUCTIONS FROM FDA TO ADDRESS A MANUFACTURER REPORT SEQUENCE NUMBER ISSUE THAT OCCURRED WHEN THIS MDR OR SUPPLEMENT WAS ORIGINALLY SUBMITTED TO THE FDA ON JAN 09 2007. EVALUATION SUMMARY: THE INFUSION PUMP WAS TESTED FOR FLOW ACCURACY AT 100 ML/HOUR, THE INFUSION PUMP FAILED THE ACCURACY TEST. THE SPECIFICATION OF THIS DEVICE IS +/-5%. THE REPORTED CONDITION OF PUMP FAILED ACCURACY TESTING WAS CONFIRMED BY THE BAXTER REPAIR TECHNICIAN. THE FAILURE WAS DETERMINED TO BE A FAULTY PUMP HEAD MODULE ASSEMBLY., WHICH WAS REPLACED. THE DEVICE WAS TESTED BY THE REPAIR TECHNICIAN AND RETURNED TO SERVICE.
THE DEVICE WAS RETURNED FOR SERVICE. DURING SERVICE TESTING, THE BAXTER TECHNICIAN REPORTED AN INFUSION PUMP WITH CONDITION OF PUMP FAILED ACCURACY TESTING. PER THE SERVICE SHOP, THE HOSPITAL REPRESENTATIVE STATED THAT THERE HAVE BEEN NO REPORTS OF ANY PATIENT INJURY OR MEDICAL INTERVENTION. NO ADDITIONAL INFORMATION IS KNOWN.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | COLLEAGUE CX VOLUMETRIC INFUSION PUMP | INFUSION PUMP | FRN | BAXTER HEALTHCARE PTE. LTD. | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |