FDA Adverse Event Malfunction Summary report: N

COLLEAGUE CX VOLUMETRIC INFUSION PUMP

MDR report key: 1183636 · Received October 4, 2008

Report

Report Number
6000001-2007-00766
Event Type
Malfunction
Date Received
October 4, 2008
Date of Event
December 14, 2006
Report Date
December 14, 2006
Manufacturer
BAXTER HEALTHCARE PTE. LTD.
Product Code
FRN
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GA, US
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 1

THIS REPORT IS BEING RESUBMITTED IN ACCORDANCE WITH INSTRUCTIONS FROM FDA TO ADDRESS A MANUFACTURER REPORT SEQUENCE NUMBER ISSUE THAT OCCURRED WHEN THIS MDR OR SUPPLEMENT WAS ORIGINALLY SUBMITTED TO THE FDA ON JAN 09 2007. EVALUATION SUMMARY: THE INFUSION PUMP WAS TESTED FOR FLOW ACCURACY AT 100ML/HOUR, THE INFUSION PUMP FAILED THE ACCURACY TEST. THE SPECIFICATION OF THIS DEVICE IS +/-5%. THE REPORTED CONDITION OF PUMP HEAD MODULE FAILED ACCURACY TESTING WITH AN UNDERINFUSION WAS CONFIRMED BY THE BAXTER REPAIR TECHNICIAN. THE FAILURE WAS DETERMINED TO BE A FAULTY PUMP HEAD MODULE ASSEMBLY, WHICH WAS REPLACED. THE DEVICE WAS TESTED BY THE REPAIR TECHNICIAN AND RETURNED TO SERVICE.

Description of Event or Problem · 1

THE DEVICE WAS RETURNED FOR SERVICE. DURING SERVICE TESTING, THE BAXTER TECHNICIAN REPORTED AN INFUSION PUMP WITH A CONDITION OF PUMP HEAD MODULE FAILED ACCURACY TESTING WITH AN UNDERINFUSION. PER THE SERVICE SHOP, THE HOSPITAL REPRESENTATIVE STATED THAT THERE HAVE BEEN NO REPORTS OF ANY PATIENT INJURY OR MEDICAL INTERVENTION. NO ADDITIONAL INFORMATION IS KNOWN.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 COLLEAGUE CX VOLUMETRIC INFUSION PUMP INFUSION PUMP FRN BAXTER HEALTHCARE PTE. LTD. NA NA

Patients

Seq Age Sex Outcome Treatment
1