FDA Adverse Event
Malfunction
Summary report: N
COLLEAGUE 3 VOLUMETRIC INFUSION PUMP
MDR report key: 1183601
·
Received October 4, 2008
Report
- Report Number
- 6000001-2007-00730
- Event Type
- Malfunction
- Date Received
- October 4, 2008
- Date of Event
- December 14, 2006
- Report Date
- December 14, 2006
- Manufacturer
- BAXTER HEALTHCARE PTE. LTD.
- Product Code
- FRN
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OH, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THIS REPORT IS BEING RESUBMITTED IN ACCORDANCE WITH INSTRUCTIONS FROM FDA TO ADDRESS A MANUFACTURER REPORT SEQUENCE NUMBER ISSUE THAT OCCURRED WHEN THIS MDR OR SUPPLEMENT WAS ORIGINALLY SUBMITTED TO THE FDA ON JAN 08 2007. EVALUATION SUMMARY: THE CONDITION OF UNDERINFUSING ON CHANNEL A WAS CONFIRMED. INSPECTION OF THE DEVICE FOUND THAT THE UNDERINFUSION WAS CAUSED BY A FAULTY PUMP HEAD MODULE. THE PUMP HEAD MODULE WAS REPLACED ON CHANNEL A.
Description of Event or Problem · 1
DURING SERVICE BY BAXTER, THE INFUSION PUMP WAS UNDERINFUSING ON CHANNEL A. ACCORDING TO THE HOSPITAL REPRESENTATIVE, NO PATIENT INJURY OR MEDICAL INTERVENTION HAD BEEN REPORTED RELATED TO THE DEVICE. NO ADDITIONAL INFORMATION OR CONTACT WAS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | COLLEAGUE 3 VOLUMETRIC INFUSION PUMP | INFUSION PUMP | FRN | BAXTER HEALTHCARE PTE. LTD. | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |