FDA Adverse Event Malfunction Summary report: N

COLLEAGUE CX VOLUMETRIC INFUSION PUMP

MDR report key: 1183594 · Received October 4, 2008

Report

Report Number
6000001-2007-01575
Event Type
Malfunction
Date Received
October 4, 2008
Date of Event
December 21, 2006
Report Date
December 21, 2006
Manufacturer
BAXTER HEALTHCARE PTE. LTD.
Product Code
FRN
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 1

THIS REPORT IS BEING RESUBMITTED IN ACCORDANCE WITH INSTRUCTIONS FROM FDA TO ADDRESS A MANUFACTURER REPORT SEQUENCE NUMBER ISSUE THAT OCCURRED WHEN THIS MDR OR SUPPLEMENT WAS ORIGINALLY SUBMITTED TO THE FDA ON JAN 16, 2007. EVALUATION SUMMARY:EVALUATION WAS PERFORMED AND THE REPORTED CONDITION OF INACCURACY WAS CONFIRMED. INSPECTION OF THE PUMP REVEALED DEFECTIVE PUMP HEAD, CAUSED THE INACCURACY. THE PUMP HEAD MODULE WAS REPLACED. THE PUMP WAS TESTED FOR PROPER OPERATION AND PUMP FOUND TO FUNCTION PROPERLY.

Description of Event or Problem · 1

DURING SERVICE BY BAXTER, THE DEVICE FAILED ACCURACY TEST WITH UNDERDELIVERY. ACCORDING TO THE HOSPITAL REPRESENTATIVE, NO PATIENT INJURY OR MEDICAL INTERVENTION HAD BEEN REPORTED RELATED TO THE DEVICE. NO ADDITIONAL INFORMATION OR CONTACT WAS AVAILABLE.