FDA Adverse Event Death Summary report: N

INFINITY CENTRAL STATION

MDR report key: 1183423 · Received October 2, 2008

Report

Report Number
1220063-2008-00030
Event Type
Death
Date Received
October 2, 2008
Date of Event
July 18, 2008
Report Date
August 12, 2008
Manufacturer
DRAGER MEDICAL SYSTEMS, INC. (IT/M)
Product Code
DSI
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
MN, US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

THE DRAGER INFINITY CENTRALSTATION INSTRUCTIONS FOR USE EXPLAINS THAT THE SOUND LEVEL OF THE INFINITY CENTRALSTATION SHOULD BE SET ACCORDING TO THE SOUND LEVEL OF THE CARE UNIT (AMBIENT NOISE). IN SECTION 1, PAGE 1-2 OF THE INSTRUCTIONS FOR USE IT STATES THAT THE INFINITY CENTRALSTATION CPU MUST BE PLACED SO THAT IT CAN BE HEARD EASILY IF THE EXTERNAL SPEAKER BECOMES NON-FUNCTIONAL. IN ADDITION, ON PAGE 4-3 OF THE INSTRUCTIONS FOR USE THERE IS A WARNING STATEMENT THAT EXPLAINS TO ALWAYS SET THE ALARM VOLUME SO IT CAN BE HEARD DURING THE BUSIEST PERIODS OF THE DAY. THE CUSTOMER HAS CONFIRMED THAT THERE WAS NO ALLEGED MACHINE MALFUNCTION. WE WILL CONTINUE TO MONITOR FOR SIMILAR REPORTS OF THIS CONDITION.

Description of Event or Problem · 1

IT WAS REPORTED THAT ALARMS WERE GENERATED AT THE ICS FOR TELEMETRY BED T5288 BEGINNING AT 5:27 PM BUT THE ALARMS WENT UNNOTICED BY THE HOSPITAL STAFF FOR AN UNDETERMINED AMOUNT OF TIME. WHEN THE STAFF RESPONDED TO THE PT, THE PT WAS EXPERIENCING CARDIAC DISTRESS AND A NEGATIVE PT OUTCOME WAS REPORTED. THE ALARM VOLUME AT THE ICS WAS REPORTEDLY SET TO 70% AND THE ICS HAS EXTERNAL AMPLIFIERS AND CEILING SPEAKERS REPORTEDLY INSTALLED BY DRAEGER TO ENHANCE AUDIBLE ALARM DETECTION. THE CUSTOMER IS REQUESTING RECOMMENDATIONS FROM DRAGER REGARDING ALARM VOLUME SETTINGS AT THE ICS AND FOR THE SUPPLEMENTAL AMPLIFIERS AND SPEAKERS TO ENHANCE ALARM DETECTION. THE CUSTOMER HAS CONFIRMED THAT THERE WAS NO ALLEGED MACHINE MALFUNCTION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 INFINITY CENTRAL STATION PATIENT MONITOR DSI DRAGER MEDICAL SYSTEMS, INC. (IT/M) DF0C86

Patients

Seq Age Sex Outcome Treatment
1 UNK