FDA Adverse Event Malfunction Summary report: N

COLLEAGUE 3 VOLUMETRIC INFUSION PUMP

MDR report key: 1183219 · Received October 3, 2008

Report

Report Number
6000001-2007-01344
Event Type
Malfunction
Date Received
October 3, 2008
Date of Event
December 1, 2006
Report Date
December 19, 2006
Manufacturer
BAXTER HEALTHCARE PTE. LTD.
Product Code
FRN
Removal / Correction Number
6000001-12/13/05-019-C
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THIS REPORT IS BEING RESUBMITTED IN ACCORDANCE WITH INSTRUCTIONS FROM FDA TO ADDRESS A MANUFACTURER REPORT SEQUENCE NUMBER ISSUE THAT OCCURRED WHEN THIS MDR OR SUPPLEMENT WAS ORIGINALLY SUBMITTED TO THE FDA ON JAN 12, 2007. EVALUATION SUMMARY:THE REPORTED CONDITION OF FAILURE CODE 812:02 WAS CONFIRMED. INSPECTION OF THE DEVICE FOUND THAT THE CAUSE OF THE FAILURE CODE WAS DUE TO WORN GEARS IN THE SHUTTLE MOTOR. THE PUMP HEAD MODULE WAS REPLACED. THE DEVICE WAS RETURNED TO THE CUSTOMER FULLY OPERATIONAL.

Description of Event or Problem · 1

THE FACILITY REPORTED DOT CUSTOMER SERVICE A PUMP WITH FAILURE CODE 812:02. THIS FAILURE CODE OCCURRED DURING BIO-MED TESTING. THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION NECESSARY ACCORDING TO THE HOSPITAL REPRESENTATIVE. NO ADDITIONAL INFORMATION IS AVAILABLE.