FDA Adverse Event Injury Summary report: N

CR 40MM GLENOSPHERE +3MM COCR

MDR report key: 11831644 · Received May 17, 2021

Report

Report Number
0001822565-2021-01344
Event Type
Injury
Date Received
May 17, 2021
Date of Event
April 21, 2021
Report Date
September 8, 2021
Manufacturer
ZIMMER BIOMET, INC.
Product Code
KWS
PMA / PMN Number
K181611
Removal / Correction Number
N/A
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 0

(B)(4). THIS FOLLOW-UP REPORT IS BEING SUBMITTED TO RELAY ADDITIONAL INFORMATION. THE FOLLOWING SECTIONS WERE UPDATED: B4; B5; D4; G3; G6; H1; H2; H3; H6. A 40MM GLENOSPHERE (LOT 64473989) WAS RETURNED FOR EVALUATION. AS RETURNED, WITNESS MARKS AND DAMAGE ARE SEEN ON THE TAPER. THE BUFFED SURFACE EXHIBITS DAMAGE / SCUFFS. REVIEW OF THE DEVICE HISTORY RECORD(S) IDENTIFIED NO DEVIATIONS OR ANOMALIES DURING MANUFACTURING RELATED TO THE ISSUE. MEDICAL RECORDS WERE NOT PROVIDED. A DEFINITIVE ROOT CAUSE CANNOT BE DETERMINED. IF ANY FURTHER INFORMATION IS FOUND WHICH WOULD CHANGE OR ALTER ANY CONCLUSIONS OR INFORMATION, A SUPPLEMENTAL WILL BE FILED ACCORDINGLY. ZIMMER BIOMET WILL CONTINUE TO MONITOR FOR TRENDS.

Description of Event or Problem · 0

NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.

Additional Manufacturer Narrative · 1

(B)(4). CONCOMITANT MEDICAL PRODUCTS: ITEM#: 118001, VERSA-DIAL/COMP TI STD TAPER; LOT#: 984380; ITEM#: 110031422, CR PROLONG 40MM BRNG +3; LOT#: 64489343; ITEM#: 110031405, MINI TRAY STD COCR +6 OFFSET; LOT#: 64675203; DEVICE EVALUATED BY MFR: PRODUCT HAS BEEN RECEIVED BY ZIMMER BIOMET AND THE INVESTIGATION IS IN PROCESS. ONCE THE INVESTIGATION HAS BEEN COMPLETED, A FOLLOW-UP MDR WILL BE SUBMITTED.

Description of Event or Problem · 1

IT WAS REPORTED THAT PATIENT UNDERWENT AN INITIAL RIGHT SHOULDER PROCEDURE ON AN UNKNOWN DATE. THE PATIENT HAD A REVISION OF THE RIGHT SHOULDER PROCEDURE DUE TO DISASSOCIATION OF COMPONENTS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
726904 CR 40MM GLENOSPHERE +3MM COCR SHOULDER PROSTHESIS/EXTREMITIES KWS ZIMMER BIOMET, INC. N/A 64473989

Patients

Seq Age Sex Outcome Treatment
1 Hospitalization| R