NEOPUFF INFANT RESUSCITATOR
Report
- Report Number
- 9611451-2021-00575
- Event Type
- Malfunction
- Date Received
- May 16, 2021
- Date of Event
- April 2, 2021
- Report Date
- April 16, 2021
- Product Code
- BTL
- PMA / PMN Number
- K892885
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WI, US
- Reporter Occupation
- 003
Narratives
(B)(4). METHOD: THE COMPLAINT RD900 NEOPUFF INFANT RESUSCITATOR WAS RECEIVED BY FISHER & PAYKEL HEALTHCARE (F&P) FOR EVALUATION. RESULTS: VISUAL INSPECTION OF THE DEVICE REVEALED THAT GLUE WAS EVIDENT IN THE BUNCHED SECTION OF THE SPRING AND OF THE MANIFOLD OF THE MANOMETER INLET PORT. CONCLUSION: IT IS POSSIBLE THAT THE TUBE MAY HAVE BEEN PULLED DUE TO EXCESSIVE FORCE. THE NEOPUFF IS A PORTABLE, REUSABLE DEVICE USED TO ASSIST IN THE DELIVERY OF RESPIRATORY BREATHS TO INFANTS UNTIL ADEQUATE SPONTANEOUS BREATHING OCCURS. BEING A PORTABLE DEVICE, THE NEOPUFF CAN BE SUSCEPTIBLE TO IMPACT DAMAGE, FOR INSTANCE WHEN ACCIDENTALLY DROPPED OR SUBJECTED TO CONSIDERABLE EXTERNAL FORCE. THE NEOPUFF IS ASSEMBLED AND 100% TESTED ON THE PRODUCTION LINE TO VERIFY THAT EACH NEOPUFF PRODUCT CONFORMS TO CRITICAL PRODUCT SPECIFICATIONS, INCLUDING THE VALVE ASSEMBLIES AND FLOW ADJUSTMENTS. THE NEOPUFF TECHNICAL MANUAL STATES THE FOLLOWING: DROPPING THE NEOPUFF / PERIVENT INFANT RESUSCITATOR OR OTHER SIMILAR FORMS OF IMPACT MAY CAUSE DAMAGE RESULTING IN INCORRECT OPERATION OF THE UNIT. IF YOU SUSPECT DAMAGE TO HAVE OCCURRED, PLEASE PERFORM CHECKS AS OUTLINED [IN THE MANUAL] BEFORE CONNECTION TO A PATIENT. IN ADDITION THE NEOPUFF USER INSTRUCTIONS STATE THAT THE USER SHOULD "CHECK MANOMETER READS ZERO WITH NO GAS FLOW" AND CHECK THE PRESSURE SETTINGS "PRIOR TO EVERY USE OF THE NEOPUFF".
(B)(4). METHOD: THE COMPLAINT RD900 NEOPUFF INFANT RESUSCITATOR WAS NOT RETURNED TO FISHER & PAYKEL HEALTHCARE (F&P) FOR EVALUATION. OUR INVESTIGATION IS THUS BASED ON THE INFORMATION AND PHOTOGRAPH PROVIDED BY THE CUSTOMER. RESULTS: VISUAL INSPECTION OF THE PROVIDED PHOTOGRAPH REVEALED THAT PVC TUBING DETACHED FROM THE MANIFOLD PORT. CONCLUSION: WITHOUT THE COMPLAINT DEVICE, WE ARE UNABLE TO DETERMINE THE CAUSE OF THE REPORTED EVENT. HOWEVER, IT IS POSSIBLE THAT THE TUBE MAY HAVE BEEN PULLED DUE TO EXCESSIVE FORCE. THE NEOPUFF IS A PORTABLE, REUSABLE DEVICE USED TO ASSIST IN THE DELIVERY OF RESPIRATORY BREATHS TO INFANTS UNTIL ADEQUATE SPONTANEOUS BREATHING OCCURS. BEING A PORTABLE DEVICE, THE NEOPUFF CAN BE SUSCEPTIBLE TO IMPACT DAMAGE, FOR INSTANCE WHEN ACCIDENTALLY DROPPED OR SUBJECTED TO CONSIDERABLE EXTERNAL FORCE. THE NEOPUFF IS ASSEMBLED AND 100% TESTED ON THE PRODUCTION LINE TO VERIFY THAT EACH NEOPUFF PRODUCT CONFORMS TO CRITICAL PRODUCT SPECIFICATIONS, INCLUDING THE VALVE ASSEMBLIES AND FLOW ADJUSTMENTS. THE NEOPUFF TECHNICAL MANUAL STATES THE FOLLOWING: - DROPPING THE NEOPUFF / PERIVENT INFANT RESUSCITATOR OR OTHER SIMILAR FORMS OF IMPACT MAY CAUSE DAMAGE RESULTING IN INCORRECT OPERATION OF THE UNIT. IF YOU SUSPECT DAMAGE TO HAVE OCCURRED, PLEASE PERFORM CHECKS AS OUTLINED [IN THE MANUAL] BEFORE CONNECTION TO A PATIENT. IN ADDITION THE NEOPUFF USER INSTRUCTIONS STATE THAT THE USER SHOULD "CHECK MANOMETER READS ZERO WITH NO GAS FLOW" AND CHECK THE PRESSURE SETTINGS "PRIOR TO EVERY USE OF THE NEOPUFF".
A DISTRIBUTOR IN WISCONSIN REPORTED, VIA A FISHER & PAYKEL HEALTHCARE (F&P) FIELD REPRESENTATIVE, THAT THE TUBING OF A RD900 NEOPUFF INFANT RESUSCITATOR WAS NOT PROPERLY ATTACHED TO THE MANOMETER PORT OF THE VALVE ASSEMBLY. THERE WAS NO PATIENT INVOLVEMENT.
A DISTRIBUTOR IN WISCONSIN REPORTED, VIA A FISHER & PAYKEL HEALTHCARE (F&P) FIELD REPRESENTATIVE, THAT THE TUBING OF A RD900 NEOPUFF INFANT RESUSCITATOR WAS NOT PROPERLY ATTACHED TO THE MANOMETER PORT OF THE VALVE ASSEMBLY. THERE WAS NO PATIENT INVOLVEMENT.
(B)(4). WE ARE CURRENTLY IN THE PROCESS OF FINALISING OUR INVESTIGATION FOR THE COMPLAINT RD900 NEOPUFF INFANT RESUSCITATOR. WE WILL PROVIDE A FOLLOW UP REPORT UPON COMPLETION OF OUR INVESTIGATION.
A DISTRIBUTOR IN (B)(6) REPORTED, VIA A FISHER & PAYKEL HEALTHCARE (F&P) FIELD REPRESENTATIVE, THAT THE TUBING OF A RD900 NEOPUFF INFANT RESUSCITATOR WAS NOT PROPERLY ATTACHED TO THE MANOMETER PORT OF THE VALVE ASSEMBLY. THERE WAS NO PATIENT INVOLVEMENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 726738 | NEOPUFF INFANT RESUSCITATOR | BTL | BTL | RD900 | 2101331074 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |