ZEPHYR ENDOBRONCHIAL VALVE (EBV)

MDR report key: 11829451 · Received May 14, 2021

Report

Report Number
3007797756-2021-00096
Event Type
Injury
Date Received
May 14, 2021
Date of Event
April 13, 2021
Report Date
May 12, 2021
Product Code
NJK
UDI-DI
00811907030324
PMA / PMN Number
P180002
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 1

PNEUMOTHORAX IS THE MOST COMMON SIDE EFFECT ASSOCIATED WITH THE ZEPHYR VALVE TREATMENT (CRINER ET AL. A MULTICENTER RANDOMIZED CONTROLLED TRIAL OF ZEPHYR ENDOBRONCHIAL VALVE TREATMENT IN HETEROGENEOUS EMPHYSEMA (LIBERATE). AM J RESPIR CRIT CARE MED. 2018; 198 (9): 1151-1164). TARGETED LOBAR DEFLATION LIKELY CAUSES INFLATION OF THE IPSILATERAL LOBE, WHICH CAN RESULT IN A TEAR OF THE ALREADY COMPROMISED PARENCHYMAL TISSUE OF THE EMPHYSEMATOUS IPSILATERAL LOBE, RESULTING IN A PNEUMOTHORAX (CRINER ET AL. A MULTICENTER RANDOMIZED CONTROLLED TRIAL OF ZEPHYR ENDOBRONCHIAL VALVE TREATMENT IN HETEROGENEOUS EMPHYSEMA (LIBERATE). AM J RESPIR CRIT CARE MED. 2018; 198 (9): 1151-1164). IN THE LIBERATE STUDY (IDE CLINICAL STUDY USED TO SUPPORT PMA P180002'S APPROVAL), 26.6% OF THE ZEPHYR VALVE SUBJECTS EXPERIENCED A PNEUMOTHORAX IN THE TREATMENT PERIOD ([LESS THAN OR EQUAL TO 45 DAYS). THESE WERE MANAGED USING STANDARD OF CARE PROCEDURES AS PER PREVIOUSLY PUBLISHED GUIDELINES (VALIPOUR, ARSCHANG, ET AL. RESPIRATION 87.6 (2014): 513-521). IN 17.4% OF THE EVENTS, THE PNEUMOTHORAX RESOLVED WITHOUT ANY ADDITIONAL INTERVENTION WITH SUBJECTS UNDER CAREFUL OBSERVATION. IN OVER HALF THE EVENTS (56.5%), THE PNEUMOTHORAX WAS MANAGED WITH A CHEST-TUBE ONLY. AN ADDITIONAL 13% OF THE EVENTS WERE MANAGED WITH A CHEST-TUBE AND THE TEMPORARY REMOVAL OF ONE OR MORE VALVES, WHILE ANOTHER 13% OF THE EVENTS WERE MANAGED WITH A CHEST-TUBE AND REMOVAL OF ALL THE IMPLANTED VALVES. UPON SUCCESSFUL RESOLUTION OF THE PNEUMOTHORAX, REMOVED VALVES CAN BE REPLACED. PATIENTS THAT EXPERIENCED A PNEUMOTHORAX EXPERIENCED CLINICAL BENEFITS OF THE ZEPHYR VALVE TREATMENT THAT WERE SIMILAR TO THE BENEFITS EXPERIENCED BY PATIENTS WHO DID NOT HAVE A PNEUMOTHORAX. THE ZEPHYR EBV SYSTEM IFU AND PULMONX TRAINING PROGRAM BOTH SPECIFICALLY REFERENCE PNEUMOTHORAX AS A KNOWN SIDE EFFECT OF THIS PROCEDURE AND THE PUBLISHED GUIDELINES (VALIPOUR, ARSCHANG, ET AL. "EXPERT STATEMENT: PNEUMOTHORAX ASSOCIATED WITH ENDOSCOPIC VALVE THERAPY FOR EMPHYSEMA-POTENTIAL MECHANISMS, TREATMENT ALGORITHM, AND CASE EXAMPLES." RESPIRATION 87.6 (2014): 513-521). THE REPORTED EVENT ALIGNS WITH THE EXPERIENCE OBSERVED IN THE LIBERATE CLINICAL STUDY AND IS AN EXPECTED SIDE EFFECT TO THE ZEPHYR VALVE TREATMENT.

Description of Event or Problem · 1

ON (B)(6) 2021, PATIENT UNDERWENT BRONCHOSCOPIC PROCEDURE AND HAD THREE ZEPHYR ENDOBRONCHIAL VALVES (5.5 EBV, 5.5-LP EBV) PLACED IN THE LEFT UPPER LOBE. PATIENT DEVELOPED A PNEUMOTHORAX POST-PROCEDURE AND A CHEST TUBE WAS PLACED ON THE SAME DAY. ON (B)(6) 2021, ONE 5.5 EBV WAS REMOVED FROM THE L4/5. ON (B)(6) 2021, THE CHEST TUBE WAS REMOVED. ON (B)(6) 2021, THE PNEUMOTHORAX RESOLVED AND THE PATIENT WAS DISCHARGED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
721442 ZEPHYR ENDOBRONCHIAL VALVE (EBV) ENDOBRONCHIAL VALVE NJK ZEPHYR 5.5 EBV 504883-V7.1 00811907030324

Patients

Seq Age Sex Outcome Treatment
1 56 YR Hospitalization| R