FDA Adverse Event Malfunction Summary report: N

GE OEC 9800

MDR report key: 1182734 · Received October 3, 2008

Report

Report Number
1720753-2008-25021
Event Type
Malfunction
Date Received
October 3, 2008
Date of Event
July 28, 2008
Report Date
August 25, 2008
Manufacturer
GE OEC MEDICAL SYSTEMS INC.
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TN, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

A GE SVC REP PERFORMED AN ON SITE INVESTIGATION. THE LEMO CONNECTOR WAS REPAIRED AND THE X-RAY TUBE TEMPERATURE SENSOR WAS REPLACED DURING THE SVC CALL. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND PUT BACK INTO SVC.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE 9800 SYSTEM HAD AN X-RAY OVER TEMPERATURE ERROR AND THE TECHNIQUE TRACKS TO MAXIMUM. ALSO, THERE ARE BLACK BARS ON THE LEFT MONITOR. NO PT INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 GE OEC 9800 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS INC. NA NA

Patients

Seq Age Sex Outcome Treatment
1