VAPR SE ELECTRODE
Report
- Report Number
- 1221934-2008-00445
- Event Type
- Malfunction
- Date Received
- September 25, 2008
- Date of Event
- September 1, 2008
- Report Date
- September 16, 2008
- Manufacturer
- DEPUY MITEK
- Product Code
- GEI
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
DEPUY MITEK TO DATE, HAS NOT YET RECEIVED THE COMPLAINT DEVICE FOR EVALUATION. HOWEVER, THE REPORTED FAILURE MODE OF THE DEVICE IS CONSISTENT WITH FAILURES PRODUCED IN A LABORATORY ENVIRONMENT BY THE APPLICATION OF EXCESSIVE MECHANICAL FORCE. ALTHOUGH IT IS NOT CONCLUSIVE WE COULD NOT IDENTIFY ANY OTHER FACTORS THAT WOULD RESULT IN SUCH A FAILURE OTHER THAN THOSE MENTIONED IN THE INSTRUCTIONS FOR USE. ALSO, THIS IS A SINGLE USE DEVICE AND IT IS NOT ESTABLISHED IF THE DEVICE IN FACT SAW ONLY ONE USAGE, IN OTHER WORDS THERE IS THE QUESTION OF THE POSSIBILITY THAT THE COMPLAINT DEVICE HAS BEEN REPROCESSED, WHICH WOULD ADD ANOTHER DIMENSION TO THE FAILURE ISSUE. THE SURGEON IS REPORTING THAT THERE WAS NO PATIENT HARM, HOWEVER, THE BROKEN FRAGMENT WAS LEFT IN THE BODY, IT IS POSSIBLE THAT PATIENT INJURY COULD POTENTIALLY OCCUR. BECAUSE OF THIS POTENTIALITY, AN MDR IS BEING FILED TO DOCUMENT THIS EVENT. WHEN AND IF THE COMPLAINT DEVICE IS RECEIVED HERE AT DEPUY MITEK, IT WILL BE SUBJECTED TO A FAILURE ROOT CAUSE ANALYSIS. IF THE ANALYSIS IDENTIFIES ANY DEFINITIVE ROOT CAUSE OTHER THAN THE ABOVE HYPOTHESIS, A FOLLOW-UP REPORT WILL BE FILED.
OUR AFFILIATE IS REPORTING THAT DURING AN ARTHROSCOPIC SHOULDER REPAIR A PORTION OF THE DISTAL END OF A SE ELECTRODE BROKE OFF INTO THE PATIENT'S JOINT SPACE. THE FRAGMENT WAS NOT ABLE TO BE RETRIEVED FROM THE BODY, HOWEVER, THE PROCEDURE WAS CONCLUDED SUCCESSFULLY WITHOUT FURTHER INCIDENT OR HARM TO THE PATIENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | VAPR SE ELECTRODE | ELECTROSURGICAL, CUTTING AND COAGULATING | GEI | DEPUY MITEK | 227301 | 0801102 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK |