ADULT INSPIRATORY HEATED BREATHING CIRCUIT
Report
- Report Number
- 9611451-2008-00552
- Event Type
- Malfunction
- Date Received
- September 25, 2008
- Date of Event
- July 10, 2008
- Report Date
- September 1, 2008
- Manufacturer
- FISHER & PAYKEL HEALTHCARE LTD.
- Product Code
- BTT
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- OTHER
Narratives
THIS IS A MALFUNCTION THAT HAS BEEN REPORTED TO FISHER & PAYKEL HEALTHCARE BY A FOREIGN DISTRIBUTOR. THE PRODUCT IS NOT SOLD IN THE USA, BUT IT IS SIMILAR TO PRODUCT WHICH IS SOLD IN THE USA. THE ELECTRICAL RESISTANCE OF THE HEATER WIRE IN THE INSPIRATORY TUBE OF THE RT106 BREATHING CIRCUIT WAS TESTED USING A MULTIMETER. THE INSPIRATORY HEATER WIRE WAS AN OPEN LOOP DUE TO A BREAK IN THE CONNECTION BETWEEN THE HEATER WIRE, AND ONE OF THE PINS THAT CRIMPS TO THE HEATER WIRE. A LOT CHECK REVEALED 1 OTHER COMPLAINT OF THIS NATURE FOR THIS LOT NUMBER. CONCLUSION: ELECTRICAL OPEN CIRCUITS IN HEATER WIRES ARE OFTEN ASSOCIATED WITH IMPROPER CRIMPING OF THE HEATER WIRE DURING PRODUCTION. ALL BREATHING CIRCUITS ARE ELECTRICALLY TESTED DURING PRODUCTION, AND THOSE THAT FAIL ARE REJECTED. THIS SUGGESTS THE HEATER WIRE BECAME AN OPEN CIRCUIT POST PRODUCTION, POSSIBLY AS A RESULT OF A WEAK CRIMP CONNECTION. FURTHER DESIGN IMPROVEMENTS TO THE EXISTING CRIMPING MACHINES ARE IN PROGRESS. OUR MONITORING AND TRENDING OF OPEN CIRCUIT HEATER WIRES IN ADULT BREATHING CIRCUITS HAS A RATE OF OCCURRENCE WORLDWIDE FOR THE LAST YEAR OF 0.0025% TO AUGUST 2008.
A HOSPITAL IN ANOTHER COUNTRY REPORTED TO OUR DISTRIBUTOR THAT A RESPIRATORY HUMIDIFIER ALARMED, WHEN HOSPITAL STAFF REPLACED THE EXISTING BREATHING CIRCUIT WITH AN ADULT INSPIRATORY HEATED BREATHING CIRCUIT. IT SEEMED THAT THIS WAS DUE TO AN OPEN CIRCUIT HEATER WIRE. NO PATIENT CONSEQUENCE WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ADULT INSPIRATORY HEATED BREATHING CIRCUIT | BTT | FISHER & PAYKEL HEALTHCARE LTD. | RT106 | 080626 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |