FDA Adverse Event Malfunction Summary report: N

PN RELION 31G X 6MM 3B TW 50CT

MDR report key: 11823952 · Received May 14, 2021

Report

Report Number
3014704491-2021-00002
Event Type
Malfunction
Date Received
May 14, 2021
Date of Event
April 15, 2021
Report Date
June 1, 2021
Manufacturer
BECTON DICKINSON MEDICAL DEVICES CO., LTD. SUZHOU
Product Code
FMI
UDI-DI
00681131386807
PMA / PMN Number
K162516
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IA, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

H6: INVESTIGATION SUMMARY: NO SAMPLES (INCLUDING PHOTOS) WERE RETURNED THEREFORE THE COMPLAINT COULD NOT BE CONFIRMED AND THE ROOT CAUSE IS UNDETERMINED. THIS IS THE 1ST COMPLAINT FOR THE REPORTED LOT NUMBER. A REVIEW OF THE MANUFACTURING RECORDS WAS PERFORMED AND NO NON-CONFORMANCES WERE RAISED IN ASSOCIATION WITH THIS TYPE OF EVENT FOR THIS LOT. CLOG TEST AND THREAD FUNCTION WAS CONDUCTED ON 7PCS RETENTION SAMPLES. NO FAILURE WAS FOUND.

Description of Event or Problem · 0

IT WAS REPORTED THAT 1 PN RELION 31G X 6MM 3B NEEDLE WAS UNABLE TO DELIVER INSULIN/MEDICATION AND WAS DIFFICULT TO OPERATE OR NOT WORKING/FUNCTIONING. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER : MATERIAL NO. 320617; BATCH NO. 0140339. THE CONSUMER REPORTED NEEDLE CLOGGED DURING PRIMING AND STATED THAT THERE IS NO INSULIN FLOW. THE CONSUMER ALSO REPORTED THAT THE NEEDLES WILL NOT ATTACH PROPERLY ONTO THE PEN. DATE OF EVENT: UNKNOWN. SAMPLES: DISCARDED.

Additional Manufacturer Narrative · 1

DATE OF EVENT: UNKNOWN. THE DATE RECEIVED BY MANUFACTURER HAS BEEN USED FOR THIS FIELD. A DEVICE EVALUATION IS PENDING BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 1

IT WAS REPORTED THAT 1 PN RELION 31G X 6MM 3B NEEDLE WAS UNABLE TO DELIVER INSULIN/MEDICATION AND WAS DIFFICULT TO OPERATE OR NOT WORKING/FUNCTIONING. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER :MATERIAL NO. 320617, BATCH NO. 0140339.THE CONSUMER REPORTED NEEDLE CLOGGED DURING PRIMING AND STATED THAT THERE IS NO INSULIN FLOW.THE CONSUMER ALSO REPORTED THAT THE NEEDLES WILL NOT ATTACH PROPERLY ONTO THE PEN. DATE OF EVENT: UNKNOWN.SAMPLES: DISCARDED,

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
725817 PN RELION 31G X 6MM 3B TW 50CT HYPODERMIC SINGLE LUMEN NEEDLE FMI BECTON DICKINSON MEDICAL DEVICES CO., LTD. SUZHOU 0140339 00681131386807

Patients

Seq Age Sex Outcome Treatment
1