DA VINCI SURGICAL SYSTEM
Report
- Report Number
- 2955842-2008-01282
- Event Type
- Other
- Date Received
- September 22, 2008
- Date of Event
- August 25, 2008
- Report Date
- September 22, 2008
- Manufacturer
- INTUITIVE SURGICAL, INC.
- Product Code
- NAY
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GA, US
- Reporter Occupation
- OTHER
Narratives
THE INVESTIGATION CONDUCTED BY THE FIELD SERVICE ENGINEER CONCLUDED THAT SYSTEM ERROR CODES #23008 AND #212 WERE ASSOCIATED WITH THE LEFT MASTER TOOL MANIPULATOR AND/OR THE INPUT OUTPUT DISTRIBUTION BOARD (IODG PCA). THE MASTER TOOL MANIPULATOR REFERS TO THE MASTER CONTROLLERS WHICH PROVIDE THE MEANS FOR THE SURGEON TO CONTROL THE INSTRUMENTS AND ENDOSCOPE INSIDE THE PATIENT FROM THE SURGEON'S SIDE CONSOLE. ONE MTM IS ASSIGNED TO THE SURGEON'S LEFT HAND (MTML) AND TO HIS RIGHT (MTMR). THE PURPOSE OF THE IODG PCA IS TO INTERFACE ALL EXTERNAL INPUT/OUTPUT TO THE INTERNAL SYSTEM ELECTRONICS. THE SYSTEM WAS REPAIRED BY REPLACING THE AFFECTED MTML AND IODG PCA. THE SYSTEM ALARM (SYSTEM GENERATED FAULT CODE) FUNCTIONED AS DESIGNED AND THERE WAS NO INJURY TO THE PATIENT. SYSTEM ERROR CODE #23008 APPEARS WHEN THE DA VINCI S SAFETY SYSTEMS DETERMINES THAT THE ANGULAR POSITION OF ONE OR MORE ROBOTIC JOINTS ON THE SPECIFIED MANIPULATOR, AS MEASURED BY THAT JOINT'S PRIMARY CONTROL SENSOR (ENCODER) AND THE SECONDARY SENSOR (POTENTIOMETER), WERE OUT OF SPECIFIED TOLERANCE FOR AGREEMENT. UPON DETERMINING THIS CONDITION, THE SAFETY SYSTEM PUT DAVINCI IN A "RECOVERABLE SAFE STATE". SYSTEM ERROR CODE #212 OCCURS WHEN THE ACTUAL VOLTAGE TO DRIVE CURRENT THROUGH THE MOTORS DEVIATES FROM THE EXPECTED VOLTAGE BY A SPECIFIED AMOUNT. SINCE THE MOTOR'S VELOCITY IS A COMPONENT OF THE EXPECTED VOLTAGE CALCULATION, THIS ERROR CAN BE CAUSED BY A FAULTY ENCODER. THE MTML AND IODG PCA WERE RETURNED TO ISI FOR FAILURE ANALYSIS INVESTIGATION. ENGINEERING EVALUATION FOUND NO ISSUES WITH THE IODG PCA, HOWEVER, DURING EVALUATION OF THE MTML, ENGINEERING FOUND THAT A MOTOR ENCODER HAD MALFUNCTIONED, THUS GENERATING THE SYSTEM ERRORS EXPERIENCED BY THE CUSTOMER. AS OF 2008, THERE HAVE BEEN NO REPORTED RECURRENCES OF THE ISSUE AT THIS HOSPITAL.
IT WAS REPORTED THAT WHILE INSERTING INSTRUMENTS INTO THE PATIENT DURING THE BEGINNING OF A DA VINCI S PROSTATECTOMY SURGICAL PROCEDURE, THE CUSTOMER EXPERIENCED SYSTEM ERROR CODES #23008 AND #212. THE SYSTEM WAS RESTARTED, HOWEVER, THE ERROR CODES RECURRED WHEN ATTEMPTING TO HOME THE SYSTEM. WITH THE ASSISTANCE OF AN ISIS TECHNICAL SUPPORT ENGINEER, THE CUSTOMER SHUT DOWN THE SYSTEM AND EXERCISED THE LEFT MASTER TOOL MANIPULATOR. THE SYSTEM WAS RESTARTED AND THE SYSTEM ERROR CODES REAPPEARED DURING HOMING. THE SURGEON DECIDED TO ABORT THE PLANNED SURGICAL PROCEDURE AND NO PATIENT HARM WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | DA VINCI SURGICAL SYSTEM | ENDOSCOPIC INSTRUMENT CONTROL SYSTEM | NAY | INTUITIVE SURGICAL, INC. | IS2000 A5.1P6 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other | DA VINCI S SURGICAL SYSTEM INSTRUMENTS| ACCESSORIES |