FDA Adverse Event Malfunction Summary report: N

FOLEY CATHETER

MDR report key: 11821206 · Received May 13, 2021

Report

Report Number
2320762-2021-00002
Event Type
Malfunction
Date Received
May 13, 2021
Date of Event
April 30, 2021
Report Date
July 23, 2021
Manufacturer
DEROYAL CIENTIFICA DE LATINOAMERICA S.R.L
Product Code
EZL
UDI-DI
00749756045773
PMA / PMN Number
K200757
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

SCAR INVESTIGATION WAS COMPLETED BY THE SUPPLIER. VISUAL EVALUATION OF THE RETURNED SAMPLE WITH A 1.5CM LONG CUT ON THE BALLOON COULD BE ATTRIBUTED TO A SHARP EDGE. NO HOLE OR PINHOLE WAS OBSERVED. IN INSTANCE THE BALLOON/CATHETER HAS COME IN CONTACT WITH SHARP EDGE. DURING INSPECTION OF INFLATION, BALLOON IS CHECKED BY INFLATING AND DEFLATING BEFORE PRINTING AND INSPECTING TEMPERATURE MEASUREMENT. BALLOON AND CATHETER SCREENING IS COMPLETED TWICE. BATCH REVIEW OF IN-PROCESS AND FINAL PACKAGING INSPECTIONS OF MANUFACTURING PROCESS WITH RESPECT TO PRODUCTION BATCH WERE REPORTED TO BE IN CONFORMANCE WITH CUSTOMER SPECIFICATIONS, AND HAVE BEEN FOUND TO BE WITHIN ALL PARAMETERS. REVIEW OF DEVICE HISTORY RECORD SHOWED NO DISCREPANCIES IN PROCESSING OF THE WORK ORDER OF THE FINISHED GOOD. SUPPLIER ROOT CAUSE WAS NOT IDENTIFIED DUE THE LACK OF EVIDENCE. DEROYAL INTERNAL BALLOON TESTING WAS COMPLETED ON 45 FOLEY CATHETER SAMPLES OF BALLOON TESTING. RESULTS SHOWED AN AVERAGE BURST VOLUME OF 38.4CC, A SAFETY FACTOR OF 7.6 TIMES THE SPECIFIED FILL VOLUME OF 5CC. DEROYAL HAS SOLD 508 CASES OF FINISHED GOOD #81-080414 IN THE LAST TWO YEARS WITH ONLY ONE SIMILAR COMPLAINT INVOLVING A LEAKAGE FROM BALLOON. DEROYAL WILL CONTINUE TO MONITOR POST-MARKET FEEDBACK AND WILL RECOGNIZE ANY FUTURE TREND DEVELOPMENT. THE INVESTIGATION IS COMPLETE AT THIS TIME. NO FURTHER INFORMATION IS AVAILABLE AT THIS TIME. WE WILL PROVIDE FOLLOW UP REPORT IF ADDITIONAL INFORMATION BECOMES AVAILABLE.

Additional Manufacturer Narrative · 1

CUSTOMER SENT A REPORT REGARDING THE MEDICAL DEVICE TO THE (B)(6), REPORT NUMBER ((B)(6)). EVALUATION OF ADVERSE EVENT REPORT DETERMINED THAT THE MEDICAL DEVICE ISSUES IS NOT REQUIRED TO BE REPORTABLE PER THE (B)(6). THE DEVICE WAS NOT RETURNED FOR EVALUATION; THEREFORE WE WERE NOT ABLE TO COMPLETE A THOROUGH INVESTIGATION. A SUPPLIER CORRECTIVE ACTION REQUEST (SCAR) HAS BEEN SENT TO VENDOR IN RELATION TO THE PRODUCT ISSUE; HOWEVER, STILL WAITING FOR COMPLETION. THE INVESTIGATION IS INCOMPLETE AT THIS TIME, A FOLLOW-UP REPORT WILL BE SUBMITTED WITH ADDITIONAL INFORMATION ONCE RECEIVED, THIS REPORT WILL BE UPDATED.

Description of Event or Problem · 1

FOLEY CATHETER HAD SLIPPED OUT AND WAS DRAINING ONTO THE FLOOR. UPON INSPECTION OF THE BALLOON CUFF, THERE WAS A 1-2 CM HOLE IN THE FOLEY CATHETER BALLOON.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
719902 FOLEY CATHETER CATHETER EZL DEROYAL CIENTIFICA DE LATINOAMERICA S.R.L 81-080414 51779966 00749756045773

Patients

Seq Age Sex Outcome Treatment
1