FDA Adverse Event Malfunction Summary report: N

MONOPOLAR CURVED SCISSORS INSTRUMENT

MDR report key: 1181973 · Received September 22, 2008

Report

Report Number
2955842-2008-01279
Event Type
Malfunction
Date Received
September 22, 2008
Report Date
September 22, 2008
Manufacturer
INTUITIVE SURGICAL, INC.
Product Code
NAY
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

RESULTS/CONCLUSIONS - THE INSTRUMENT WAS RETURNED AND EVALUATED. PER THE CUSTOMER REPORTED COMPLAINT, ENGINEERING CONFIRMED THE DISTAL END OF MAIN TUBE HAS A 1.25" LONG SECTION WITH MATERIAL REMOVED ON ONE SIDE OF THE TUBE. THE DAMAGED AREA IS PARALLEL TO TUBE AXIS AND HAS A ROUGH SURFACE FINISH. BASED ON THE LOCATION AND APPEARANCE, THE DAMAGE WAS MOST LIKELY CAUSED BY A CANNULA ACCESSORY. NO OTHER DAMAGE FOUND. ADDITIONAL INVESTIGATION IS UNDERWAY REGARDING THE CANNULA ACCESSORIES. A FOLLOW-UP MDR WILL BE SUBMITTED IF ADDITIONAL INFORMATION IS RECEIVED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE MONOPOLAR CURVED SCISSORS INSTRUMENT IS SPLINTERING. NO ADDITIONAL INFORMATION WAS PROVIDED. NO PATIENT HARM, ADVERSE OUTCOME OR INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 MONOPOLAR CURVED SCISSORS INSTRUMENT ELECTROSURGICAL INSTRUMENT NAY INTUITIVE SURGICAL, INC. 420179-08 0803141 311

Patients

Seq Age Sex Outcome Treatment
1 DA VINCI S SURGICAL SYSTEM| ELECTROSURGICAL UNIT| ACCESSORIES