FDA Adverse Event
Malfunction
Summary report: N
MONOPOLAR CURVED SCISSORS INSTRUMENT
MDR report key: 1181973
·
Received September 22, 2008
Report
- Report Number
- 2955842-2008-01279
- Event Type
- Malfunction
- Date Received
- September 22, 2008
- Report Date
- September 22, 2008
- Manufacturer
- INTUITIVE SURGICAL, INC.
- Product Code
- NAY
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- UNKNOWN
Narratives
Additional Manufacturer Narrative · 1
RESULTS/CONCLUSIONS - THE INSTRUMENT WAS RETURNED AND EVALUATED. PER THE CUSTOMER REPORTED COMPLAINT, ENGINEERING CONFIRMED THE DISTAL END OF MAIN TUBE HAS A 1.25" LONG SECTION WITH MATERIAL REMOVED ON ONE SIDE OF THE TUBE. THE DAMAGED AREA IS PARALLEL TO TUBE AXIS AND HAS A ROUGH SURFACE FINISH. BASED ON THE LOCATION AND APPEARANCE, THE DAMAGE WAS MOST LIKELY CAUSED BY A CANNULA ACCESSORY. NO OTHER DAMAGE FOUND. ADDITIONAL INVESTIGATION IS UNDERWAY REGARDING THE CANNULA ACCESSORIES. A FOLLOW-UP MDR WILL BE SUBMITTED IF ADDITIONAL INFORMATION IS RECEIVED.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE MONOPOLAR CURVED SCISSORS INSTRUMENT IS SPLINTERING. NO ADDITIONAL INFORMATION WAS PROVIDED. NO PATIENT HARM, ADVERSE OUTCOME OR INJURY WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | MONOPOLAR CURVED SCISSORS INSTRUMENT | ELECTROSURGICAL INSTRUMENT | NAY | INTUITIVE SURGICAL, INC. | 420179-08 | 0803141 311 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | DA VINCI S SURGICAL SYSTEM| ELECTROSURGICAL UNIT| ACCESSORIES |