FDA Adverse Event
Malfunction
Summary report: N
AFFINITY BED
MDR report key: 1181957
·
Received September 26, 2008
Report
- Report Number
- 1824206-2008-03610
- Event Type
- Malfunction
- Date Received
- September 26, 2008
- Date of Event
- September 17, 2008
- Report Date
- September 17, 2008
- Manufacturer
- HILL-ROM RITTER
- Product Code
- HDD
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MI, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE HILL-ROM TECHNICIAN FOUND THE HI/LOW DRIVE SCREW WAS DIRTY CAUSING THE HI/LOW FUNCTION OF THE BED TO DRIFT. THE TECHNICIAN CLEANED THE DRIVE TO REPAIR THE BED. CHAPTER 6 OF THE SERVICE MANUAL DOCUMENTS A FUNCTION CHECK FOR HI/LOW DRIVE SCREW AS PART OF PREVENTIVE MAINTENANCE. AS PART OF THE PROCEDURE, THIS SHOULD BE INSPECTED FOR PROPER LUBRICATION.
Description of Event or Problem · 1
THE HI/LOW FUNCTION OF THE BED DRIFTS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | AFFINITY BED | BIRTHING BED | HDD | HILL-ROM RITTER | 3600 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |