FDA Adverse Event Malfunction Summary report: N

CAMERA HEAD AC - C-MOUNT

MDR report key: 11819543 · Received May 13, 2021

Report

Report Number
1221934-2021-01533
Event Type
Malfunction
Date Received
May 13, 2021
Date of Event
April 9, 2021
Report Date
May 12, 2021
Manufacturer
MEDOS INTERNATIONAL SARL
Product Code
FWF
UDI-DI
10886705028733
PMA / PMN Number
EXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

ADDITIONAL NARRATIVE: UDI: (B)(4). INVESTIGATION SUMMARY: THE COMPLAINT DEVICE WAS RECEIVED AT THE SERVICE CENTER AND EVALUATED. IT WAS REPORTED THAT THE SHEATH WAS NOT FITTING. PER SERVICE REPORTS, THIS COMPLAINT CAN BE CONFIRMED. DURING THE SERVICE EVALUATION THE FOLLOWING DEFECTS WERE IDENTIFIED: INTERMITTENT / NO VIDEO. CORD DAMAGED. MINOR SCRATCHES ON THE DEVICE. THE DEVICE WAS HOWEVER DEEMED NON-REPAIRABLE AND WAS BEING PLACED INTO LONG TERM HOLD. THE FAULTY PARTS WAS IDENTIFIED AS THE ROOT CAUSE FOR THE DEVICE FAILURE DURING THE SERVICE EVALUATION. MANUFACTURING RECORD EVALUATION IS NOT REQUIRED AS THE REPORTED EVENT IS NOT ASSOCIATED WITH THE MANUFACTURING PROCESS AND/OR THE POTENTIAL CAUSE OF THE DEFECT CANNOT BE ASSOCIATED TO MANUFACTURING. AT THIS POINT IN TIME, NO CORRECTIVE ACTION IS REQUIRED, AND NO FURTHER ACTION IS WARRANTED. DEPUY MITEK WILL CONTINUE TO TRACK ANY RELATED COMPLAINTS WITHIN THIS DEVICE FAMILY AS A MEANS OF MONITORING THE EXTENT WITH WHICH THIS COMPLAINT IS OBSERVED IN THE FIELD.

Description of Event or Problem · 1

IT WAS REPORTED BY SALES REP THAT DURING AN UNKNOWN PROCEDURE ON (B)(6) 2021, IT WAS OBSERVED THAT THE CORD ON THE CAMERA HEAD AC - C-MOUNT DEVICE WAS LOOSE. DURING IN-HOUSE ENGINEERING EVALUATION, IT WAS OBSERVED THAT THE DEVICE WAS WORKING INTERMITTENTLY WITH NO VIDEO. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
717228 CAMERA HEAD AC - C-MOUNT ENDOSCOPIC VIDEO CAMERA FWF MEDOS INTERNATIONAL SARL 242401 10886705028733

Patients

Seq Age Sex Outcome Treatment
1