FDA Adverse Event Malfunction Summary report: N

ADVANTA

MDR report key: 1181909 · Received September 25, 2008

Report

Report Number
1824206-2008-03597
Event Type
Malfunction
Date Received
September 25, 2008
Date of Event
August 27, 2008
Report Date
August 27, 2008
Manufacturer
HILL-ROM RITTER
Product Code
FNL
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER

Narratives

Description of Event or Problem · 1

HILL-TECH FOUND BED IN REPAIR AREA AND THE BED WOULD NOT GO INTO REVERSE TRENDELENBURG. NO INJURY REPORTED. TECH FOUND THE WIRING FROM HEAD HI/LO MOTOR WAS NOT FULLY SEATED AT LOGIC BOARD. RESEATED THE PINS FROM HEAD HI/LO MOTOR TO RESOLVE THE ISSUE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ADVANTA AC POWERED HOSPITAL BED FNL HILL-ROM RITTER 1600 NA

Patients

Seq Age Sex Outcome Treatment
1 UNK