FDA Adverse Event
Malfunction
Summary report: N
ADVANTA
MDR report key: 1181909
·
Received September 25, 2008
Report
- Report Number
- 1824206-2008-03597
- Event Type
- Malfunction
- Date Received
- September 25, 2008
- Date of Event
- August 27, 2008
- Report Date
- August 27, 2008
- Manufacturer
- HILL-ROM RITTER
- Product Code
- FNL
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- OTHER
Narratives
Description of Event or Problem · 1
HILL-TECH FOUND BED IN REPAIR AREA AND THE BED WOULD NOT GO INTO REVERSE TRENDELENBURG. NO INJURY REPORTED. TECH FOUND THE WIRING FROM HEAD HI/LO MOTOR WAS NOT FULLY SEATED AT LOGIC BOARD. RESEATED THE PINS FROM HEAD HI/LO MOTOR TO RESOLVE THE ISSUE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ADVANTA | AC POWERED HOSPITAL BED | FNL | HILL-ROM RITTER | 1600 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK |