FDA Adverse Event Malfunction Summary report: N

GE OEC 9600

MDR report key: 1181886 · Received September 19, 2008

Report

Report Number
1720753-2008-26001
Event Type
Malfunction
Date Received
September 19, 2008
Date of Event
August 28, 2008
Report Date
September 18, 2008
Manufacturer
GE OEC MEDICAL SYSTEMS INC
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AR, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REPRESENTATIVE PERFORMED AN ON SITE INVESTIGATION. THE BATTERIES WERE REPLACED DURING THE SERVICE CALL. THE SYSTEM WAS TESTED, AND FOUND TO BE WORKING AS INTENDED, AND PUT BACK INTO SERVICE.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE 9600 SYSTEM HAD A CHARGER ERROR DURING A CASE. THE 9600 SYSTEM HAD TO BE REBOOTED AND THE PROCEDURE WAS COMPLETED WITHOUT FURTHER INCIDENT. NO PATIENT INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 GE OEC 9600 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS INC NA NA

Patients

Seq Age Sex Outcome Treatment
1