FDA Adverse Event
Malfunction
Summary report: N
GE OEC 9800
MDR report key: 1181852
·
Received September 19, 2008
Report
- Report Number
- 1720753-2008-26070
- Event Type
- Malfunction
- Date Received
- September 19, 2008
- Date of Event
- August 28, 2008
- Report Date
- September 19, 2008
- Manufacturer
- GE OEC MEDICAL SYSTEMS INC.
- Product Code
- JAA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
Additional Manufacturer Narrative · 1
A GE SERVICE REPRESENTATIVE PERFORMED AN ON SITE INVESTIGATION. THE SNUBBER BOARD, X-RAY TUBE, AND HIGH TENSION TANK WERE REPLACED. THE COLLIMATOR AND FILAMENT WERE CALIBRATED DURING THE SERVICE CALL. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND PUT BACK INTO SERVICE.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE 9800 SYSTEM MADE A CRACKLING NOISE WHEN IMAGING. NO PATIENT INJURY WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | GE OEC 9800 | FLUOROSCOPIC X-RAY | JAA | GE OEC MEDICAL SYSTEMS INC. | NA | NA |