FDA Adverse Event
Malfunction
Summary report: N
GE OEC 9900
MDR report key: 1181823
·
Received September 19, 2008
Report
- Report Number
- 1720753-2008-26061
- Event Type
- Malfunction
- Date Received
- September 19, 2008
- Date of Event
- August 27, 2008
- Report Date
- September 19, 2008
- Manufacturer
- GE OEC MEDICAL SYSTEMS INC.
- Product Code
- JAA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
Additional Manufacturer Narrative · 1
A GE SERVICE REP PERFORMED AN ON SITE INVESTIGATION. THE FAILURE COULD NOT BE DUPLICATED. THE HIGH VOLTAGE CABLE WAS CLEANED AND RE-GREASED. THE FILAMENT WAS CALIBRATED DURING THE SERVICE CALL. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND PUT BACK INTO SERVICE.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE X-RAY TUBE ON THE 9900 SYSTEM WAS ARCING. NO PATIENT INJURY WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | GE OEC 9900 | FLUOROSCOPIC X-RAY | JAA | GE OEC MEDICAL SYSTEMS INC. | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |