LIGAMAX-5MM ENDO CLIP APPLIER
Report
- Report Number
- 3005075853-2008-01995
- Event Type
- Malfunction
- Date Received
- September 30, 2008
- Date of Event
- August 19, 2008
- Report Date
- August 19, 2008
- Manufacturer
- ETHICON ENDO-SURGERY, LLC
- Product Code
- FZP
- PMA / PMN Number
- K050344
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FR
- Reporter Occupation
- OTHER
Narratives
EMPTY. EVALUATION SUMMARY - THE ANALYSIS RESULTS FOUND THAT THE EL5ML DEVICE WAS RECEIVED IN GOOD VISUAL CONDITION. WHEN TESTING THE DEVICE FOR FUNCTIONALITY, IT WAS NOTED TO BE EMPTY AND THE LOCKOUT WAS FIRED THROUGH. AS THE IFU STATES: "DO NOT ATTEMPT TO FIRE THROUGH THE LOCKOUT. IF THE TRIGGER IS FORCED CLOSED ONCE THE LAST CLIP LOCKOUT HAS ENGAGED, THE JAWS MAY REMAIN CLOSED". WE HAVE DOCUMENTED THE CIRCUMSTANCES AS THEY WERE REPORTED TO US. IN ADDITION, COMPLAINT INFORMATION IS TRENDED ON A REGULAR BASIS TO DETERMINE IF FURTHER INVESTIGATION IS WARRANTED. A BATCH RECORD REVIEW WAS PERFORMED AND NO ANOMALIES WERE FOUND DURING THE MANUFACTURING PROCESS.
IT WAS REPORTED THAT DURING A NEPHRECTOMY PROCEDURE, THE CLIPS WOULD NOT ADVANCE. SEVERAL CLIPS REMAIN BLOCKED UNDER THE DEVICE. ANOTHER DEVICE WAS USED TO COMPLETE THE CASE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | LIGAMAX-5MM ENDO CLIP APPLIER | FZP | ETHICON ENDO-SURGERY, LLC | NA | E4KR6D |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |