FDA Adverse Event Malfunction Summary report: N

LIGAMAX-5MM ENDO CLIP APPLIER

MDR report key: 1181735 · Received September 30, 2008

Report

Report Number
3005075853-2008-01997
Event Type
Malfunction
Date Received
September 30, 2008
Report Date
August 20, 2008
Manufacturer
ETHICON ENDO-SURGERY, LLC
Product Code
FZP
PMA / PMN Number
K050344
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MN, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

EMPTY.EVALUATION SUMMARY - THE ANALYSIS RESULTS FOUND THAT THE EL5ML DEVICE WAS RECEIVED IN GOOD VISUAL CONDITION. WHEN TESTING THE DEVICE FOR FUNCTIONALITY, IT WAS NOTED TO BE EMPTY AND THE LOCKOUT WAS FIRED THROUGH. AS THE IFU STATES: "DO NOT ATTEMPT TO FIRE THROUGH THE LOCKOUT. IF THE TRIGGER IS FORCED CLOSED ONCE THE LAST CLIP LOCKOUT HAS ENGAGED, THE JAWS MAY REMAIN CLOSED." WE HAVE DOCUMENTED THE CIRCUMSTANCES AS THEY WERE REPORTED TO US. IN ADDITION, COMPLAINT INFORMATION IS TRENDED ON A REGULAR BASIS TO DETERMINE IF FURTHER INVESTIGATION IS WARRANTED. A BATCH REVIEW WAS PERFORMED AND NO ANOMALIES WERE FOUND DURING THE MANUFACTURING PROCESS.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE, THE DEVICE WOULD NOT RELEASE. NO ADDITIONAL INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 LIGAMAX-5MM ENDO CLIP APPLIER FZP ETHICON ENDO-SURGERY, LLC NA E4LE80

Patients

Seq Age Sex Outcome Treatment
1