LIGAMAX-5MM ENDO CLIP APPLIER
Report
- Report Number
- 3005075853-2008-01997
- Event Type
- Malfunction
- Date Received
- September 30, 2008
- Report Date
- August 20, 2008
- Manufacturer
- ETHICON ENDO-SURGERY, LLC
- Product Code
- FZP
- PMA / PMN Number
- K050344
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MN, US
- Reporter Occupation
- OTHER
Narratives
EMPTY.EVALUATION SUMMARY - THE ANALYSIS RESULTS FOUND THAT THE EL5ML DEVICE WAS RECEIVED IN GOOD VISUAL CONDITION. WHEN TESTING THE DEVICE FOR FUNCTIONALITY, IT WAS NOTED TO BE EMPTY AND THE LOCKOUT WAS FIRED THROUGH. AS THE IFU STATES: "DO NOT ATTEMPT TO FIRE THROUGH THE LOCKOUT. IF THE TRIGGER IS FORCED CLOSED ONCE THE LAST CLIP LOCKOUT HAS ENGAGED, THE JAWS MAY REMAIN CLOSED." WE HAVE DOCUMENTED THE CIRCUMSTANCES AS THEY WERE REPORTED TO US. IN ADDITION, COMPLAINT INFORMATION IS TRENDED ON A REGULAR BASIS TO DETERMINE IF FURTHER INVESTIGATION IS WARRANTED. A BATCH REVIEW WAS PERFORMED AND NO ANOMALIES WERE FOUND DURING THE MANUFACTURING PROCESS.
IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE, THE DEVICE WOULD NOT RELEASE. NO ADDITIONAL INFORMATION WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | LIGAMAX-5MM ENDO CLIP APPLIER | FZP | ETHICON ENDO-SURGERY, LLC | NA | E4LE80 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |