FDA Adverse Event Injury Summary report: N

IMPL TAPERED SCR-V SBM 4. 7MM 4.5MM 10MM

MDR report key: 11816432 · Received May 13, 2021

Report

Report Number
0002023141-2021-01245
Event Type
Injury
Date Received
May 13, 2021
Date of Event
April 6, 2021
Report Date
August 19, 2021
Manufacturer
ZIMMER DENTAL
Product Code
DZE
UDI-DI
00889024020047
PMA / PMN Number
K061410
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
DENTIST

Narratives

Additional Manufacturer Narrative · 0

ONE IMPL TAPERED SCR-V SBM 4. 7MM 4.5MM 10MM (TSVWB10) WAS RETURNED FOR INVESTIGATION. VISUAL INSPECTION OF THE AS RETURNED PRODUCT IDENTIFIED WORN MARKINGS DUE TO USAGE. NO DAMAGE. THE RETURNED DEVICE WAS MEASURED AND VERIFIED TO MATCH DHR SPECIFICATIONS DRAWING. FUNCTIONAL TESTING TO RECREATE THE REPORTED EVENT COULD NOT BE PERFORMED DUE TO THE NATURE OF THE DEVICE AND EVENT. NO PRE-EXISTING CONDITIONS WERE NOTED ON THE PER. THE REPORTED DEVICES LOCATION IS #30 AND USAGE ON THE SAME DAY. PICTURES OR X-RAY IMAGES WERE NOT PROVIDED. APPROPRIATE DOCUMENTATION WAS REVIEWED. DHR REVIEW WAS COMPLETED FOR THE SUBJECT LOT NUMBER (1242120). IT WAS CONFIRMED THAT ALL OPERATIONS AND INSPECTIONS WERE EXECUTED AS PER APPLICABLE PROCEDURE. NO DEVIATIONS OR NON-CONFORMANCES, WHICH COULD HAVE CAUSED OR CONTRIBUTED TO THE REPORTED EVENT WAS NOTED AS PART OF THE DHR. LOT WAS INSPECTED AND PASSED ALL ACCEPTANCE CRITERIA BY QA. COMPLAINT HISTORY REVIEW WAS PERFORMED FOR THE REPORTED LOT NUMBER (1242120) FOR SIMILAR EVENT AND NO OTHER COMPLAINT WAS IDENTIFIED. BASED ON THE AVAILABLE INFORMATION, DEVICE MALFUNCTION DID NOT OCCUR AND THE REPORTED EVENT WAS NON-VERIFIABLE.

Description of Event or Problem · 0

NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.

Additional Manufacturer Narrative · 1

ZIMMER BIOMET (B)(4). AGE AND DATE OF BIRTH UNKNOWN / NOT PROVIDED. PATIENT WEIGHT UNKNOWN / NOT PROVIDED. PMA/510K: K011028, K013227. A SUMMARY INVESTIGATION HAS BEEN COMPLETED FOR LACK OF PRIMARY STABILITY EVENTS THAT DO NOT ALLEGE A DEFICIENCY WITH THE IMPLANT AND IDENTIFIED THAT THE REPORTED EVENT IS LIKELY DUE TO BIOLOGICAL FACTORS WHICH HAVE AN ADVERSE EFFECT ON IMPLANT STABILITY OR IS RELATED TO SURGICAL TECHNIQUE AND INSERTION TORQUE. DUE TO A WIDE RANGE OF EXTERNAL (NON-DESIGN/ NON-MANUFACTURING RELATED), IDENTIFYING A DEFINITIVE ROOT CAUSE IS GENERALLY NOT POSSIBLE. SHOULD ADDITIONAL INFORMATION BE RECEIVED WHICH INDICATES THAT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT, AN ADDITIONAL REPORT WILL BE SUBMITTED. SUMMARY REPORTING.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE FIRST IMPLANT WAS UNABLE TO BE PLACED DUE TO BEING TOO WIDE. DR. ATTEMPTED TO PLACE ANOTHER IMPLANT AT THE SAME SITE BUT HEARD A CROCK/POP SOUND UPON TORQUEING. PATIENT WILL NOT RETURN FOR ADDITIONAL APPOINTMENTS. TOOTH #30.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
717927 IMPL TAPERED SCR-V SBM 4. 7MM 4.5MM 10MM DENTAL IMPLANT DZE ZIMMER DENTAL TSVWB10 1242120 00889024020047

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention