FDA Adverse Event Malfunction Summary report: N

PROGRAMMING SOFTWARE

MDR report key: 1181593 · Received September 25, 2008

Report

Report Number
1644487-2008-02346
Event Type
Malfunction
Date Received
September 25, 2008
Date of Event
September 2, 2008
Report Date
September 2, 2008
Manufacturer
CYBERONICS, INC.
Product Code
LYJ
PMA / PMN Number
P970003
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
OTHER

Narratives

Description of Event or Problem · 1

INITIAL REPORTER INDICATED THAT THEIR HANDHELD WAS NOT ABLE TO COMMUNICATE WITH THE PATIENT'S GENERATOR AND WAS FREEZING UPON INTERROGATION. THE HANDHELD CONTAINED 7.1 PROGRAMMING SOFTWARE. THE HANDHELD WAS UNPLUGGED AND THE AFTER A FLASHCARD REINSERTION WITH SOFT RESET THE PATIENT WAS REINTERROGATED SUCCESSFULLY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PROGRAMMING SOFTWARE LYJ CYBERONICS, INC. MODEL 250 523496

Patients

Seq Age Sex Outcome Treatment
1