TAXUS EXPRESS2 PACLITAXEL-ELUTING CORONARY STENT SYSTEM
Report
- Report Number
- 2134265-2008-02808
- Event Type
- Malfunction
- Date Received
- September 25, 2008
- Date of Event
- September 3, 2008
- Report Date
- September 3, 2008
- Manufacturer
- BOSTON SCIENTIFIC
- Product Code
- NIQ
- PMA / PMN Number
- P030025
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- PHYSICIAN
Narratives
AS THE UNIT HAS NOT BEEN RETURNED, THE COMPLAINT INVESTIGATION SITE COULD NOT PERFORM A TECHNICAL ANALYSIS. A REVIEW OF THE MFG DOCUMENTATION FOR THIS PARTICULAR BATCH FOUND THAT THE DEVICE MET ITS MATERIAL, ASSEMBLY AND PRODUCT SPECIFICATIONS AT THE TIME OF RELEASE TO DISTRIBUTION. THE MOST PROBABLE ROOT CAUSE CLASSIFICATION IS CONSIDERED OPERATIONAL CONTEXT DUE TO ANATOMICAL/PROCEDURAL FACTORS ENCOUNTERED DURING THE PROCEDURE THE PERFORMANCE OF THE DEVICE WAS LIMITED.
IT WAS REPORTED THAT DURING A CORONARY ARTERY DRUG ELUTING STENTING TREATMENT PROCEDURE THERE WAS STENT DAMAGE. THE LESION BEING TREATED WAS 99% STENOSED WITH CALCIFICATION IN AN EXTREMELY TORTUOUS PROXIMAL LEFT CIRCUMFLEX TO THE POSTEROLATERAL BRANCH. A NON BSC GUIDE CATHETER DID NOT PROVIDE GOOD BACK UP AND IT WAS CHANGED TO A DIFFERENT NON BSC GUIDE CATHETER. IVUS THEN CROSSED THE LESION. FIRST, A TAXUS 3.5X24MM WAS IMPLANTED AT LCX PL FOR INSTANT RESTENOSIS OF A PREVIOUSLY IMPLANTED NON BSC STENT. THE PROXIMAL LEFT CIRCUMFLEX WAS DILATED WITH A QUANTUM MAVERICK 3.5X8MM. AN ATTEMPT WAS MADE TO IMPLANT A TAXUS 3.5X28MM, HOWEVER, IT COULD NOT CROSS THE LESION WITH THE SEVERE CALCIFICATION. SEVERAL ATTEMPTS WERE MADE AND IT WAS THEN NOTED THAT THE PROXIMAL EDGE OF THE STENT LIFTED UP. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. PATIENT STATUS IS GOOD WITH NO COMPLICATIONS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | TAXUS EXPRESS2 PACLITAXEL-ELUTING CORONARY STENT SYSTEM | NIQ STENT, CORONARY, DRUG-ELUTING | NIQ | BOSTON SCIENTIFIC | 3.50X28MM | 0011487134 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | GC: BRITE TIP |