FDA Adverse Event Injury Summary report: N

RESPIRONICS

MDR report key: 11813908 · Received May 12, 2021

Report

Report Number
2031642-2021-03696
Event Type
Injury
Date Received
May 12, 2021
Date of Event
April 13, 2021
Manufacturer
RESPIRONICS CALIFORNIA, LLC
Product Code
MNT
UDI-DI
00884838020054
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NV, US
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

THE ALLEGED MALFUNCTIONED BLOWER MOTOR WAS RECEIVED BY PHILIPS FAILURE INVESTIGATIONS AND INSPECTED/EVALUATED. UPON FURTHER EVALUATION, IT WAS DETERMINED THAT THE CUSTOMER COMPLAINT COULD NOT BE VERIFIED AND THAT THE RETURNED BLOWER MOTOR PASSED ALL TESTING WITH NO FAULT FOUND.

Description of Event or Problem · 1

IT WAS REPORTED TO PHILIPS THAT WHILE DELIVERING THERAPY TO A PATIENT, THE RESPIRONICS V60 VENTILATOR DEVICE GENERATED A BLOWER TEMPERATURE TOO HIGH ALARM, OVER HEATED, AND ENTERED INTO AN INOPERABLE STATE. THE CUSTOMER REPORTED THAT THE UNIT WAS IN USE ON A PATIENT AT THE TIME OF THE REPORTED DEVICE SYMPTOM AND ADVERSE EVENT. A PHILIPS FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED TO THE CUSTOMER SITE AND DETERMINED THAT THE BLOWER REQUIRED REPLACEMENT. NO DIAGNOSTIC REPORT (DRPT) WAS PROVIDED FOR REVIEW. THE FSE REPLACED THE BLOWER ASSEMBLY TO RESOLVE THE ISSUE. THE DEVICE PASSED PERFORMANCE VERIFICATION TESTING AFTER REPAIR WAS COMPLETED AND WAS RETURNED TO SERVICE. THIS REPORTER STATED THAT A PATIENT OF UNKNOWN AGE, GENDER, HEIGHT, AND WEIGHT WAS ADMITTED TO A HOSPITAL ON AN UNKNOWN DATE WITH AN ADMITTING DIAGNOSIS OF CORONAVIRUS (COVID19). RELEVANT MEDICAL HISTORY INCLUDED MULTIPLE CO-MORBIDITIES. NO RELEVANT PAST DRUG HISTORY OR RELEVANT CONCOMITANT MEDICAL PRODUCTS WERE REPORTED. WHILE ADMITTED ON AN UNKNOWN DATE, THE PATIENT WAS PRESCRIBED AVERAGE VOLUME-ASSURED PRESSURE SUPPORT (AVAPS) THERAPY VIA THE RESPIRONICS V60 VENTILATOR; PRESCRIPTION, DEVICE SETTINGS, CONFIGURATION, PATIENT CIRCUIT, AND PATIENT INTERFACE NOT REPORTED. WHILE ADMITTED ON (B)(6) 2021, THE PATIENT WAS RECEIVING THERAPY VIA THE V60 DEVICE, THE DEVICE GENERATED A BLOWER TEMPERATURE TOO HIGH ALARM, SHUT DOWN, THE PATIENT EXPERIENCED A CARDIOPULMONARY ARREST, CARDIOPULMONARY RESUSCITATION EFFORTS WERE INITIATED, THE PATIENT EXPERIENCED A RETURN OF SPONTANEOUS CIRCULATION, WAS THEN INTUBATED AND PLACED ON MECHANICAL VENTILATION; BRAND, MODEL, AND PRESCRIPTION NOT REPORTED. DAYS LATER ON AN UNKNOWN DATE, THE PATIENT EXPERIENCED AN OUTCOME OF DEATH DUE TO THEIR MULTIPLE CO-MORBIDITIES. THE CAUSE OF DEATH WAS NOT REPORTED. THERE WAS NO ALLEGATION AGAINST THE V60 VENTILATOR AND THE PATIENT¿S OUTCOME. THERE WERE NO CURTAINS, DRAPES, OR OTHER MATERIALS ON THE V60 VENTILATOR AT THE TIME OF THE DEVICE BEHAVIOR. NO RELEVANT LABORATORY DATA WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
710228 RESPIRONICS VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE MNT RESPIRONICS CALIFORNIA, LLC V60 00884838020054

Patients

Seq Age Sex Outcome Treatment
1 Unknown Required Intervention