MELOLIN STERILE 20X10CM CTN 100

MDR report key: 11813820 · Received May 12, 2021

Report

Report Number
8043484-2021-01090
Event Type
Malfunction
Date Received
May 12, 2021
Date of Event
April 21, 2021
Report Date
September 1, 2021
Manufacturer
SMITH & NEPHEW MEDICAL LTD.
Product Code
NAC
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SI
Reporter Occupation
OTHER

Narratives

Description of Event or Problem · 0

IT WAS REPORTED THAT A MELOLIN STERILE 20X10CM CTN 100 PACKAGE, CAN NOT BE OPENED ASEPTICALLY, BECAUSE THE STERILE PART IS NOT SUFFICIENTLY INTEGRAL AND STRONG AND TEARS WHEN OPENED. NO CASE INVOLVED.

Additional Manufacturer Narrative · 1

SMITH & NEPHEW, INC. IS SUBMITTING THIS REPORT PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED UPON INFORMATION WHICH SMITH & NEPHEW, INC. HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY FDA, SMITH & NEPHEW, INC., OR ITS EMPLOYEES, THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, SMITH & NEPHEW, INC., OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT.

Additional Manufacturer Narrative · 1

THE DEVICE INTENDED TO BE USED IN TREATMENT HAS BEEN RETURNED AND EVALUATED ESTABLISHING A RELATIONSHIP WITH THE REPORTED EVENT. THE VISUAL INSPECTION REPORTED NO ISSUES, FUNCTIONAL ASSESSMENT CONFIRMED THEY WERE DIFFICULT TO OPEN. THE ROOT CAUSE WAS IDENTIFIED AS A GLUE FIRMNESS AND STORAGE ENVIRONMENT ISSUE. THE MANUFACTURING RECORDS SHOW NO EVIDENCE THAT THE PRODUCT DID NOT MEET SPECIFICATION AT THE TIME OF MANUFACTURE. THE COMPLAINT HISTORY FILE CONTAINS FURTHER INSTANCES. THIS INVESTIGATION IS NOW COMPLETE WITH NO FURTHER ACTION DEEMED NECESSARY. SMITH + NEPHEW WILL CONTINUE TO MONITOR FOR ANY ADVERSE TRENDS RELATING TO THIS PRODUCT RANGE.

Description of Event or Problem · 1

IT WAS REPORTED THAT A MELOLIN STERILE 20X10CM CTN 100 PACKAGE, CAN NOT BE OPENED ASEPTICALLY. NO CASE INVOLVED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
712113 MELOLIN STERILE 20X10CM CTN 100 MELOLIN STERILE 20X10CM CTN 100 NAC SMITH & NEPHEW MEDICAL LTD. 2032B

Patients

Seq Age Sex Outcome Treatment
1