PULSE GEN MODEL 101

MDR report key: 1181372 · Received September 23, 2008

Report

Report Number
1644487-2008-02296
Event Type
Malfunction
Date Received
September 23, 2008
Date of Event
July 26, 2001
Report Date
September 4, 2008
Manufacturer
CYBERONICS, INC.
Product Code
LYJ
PMA / PMN Number
P970003
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
NOT APPLICABLE

Narratives

Description of Event or Problem · 1

THE REPORTER INDICATED THAT IT WAS DURING THE ANALYSIS OF A VNS PATIENT'S GENERATOR, WHO UNDERWENT REVISION SURGERY DUE TO THE EXPECTED END OF SERVICE OF THE DEVICE, IT WAS REVEALED THAT A TRANSISTOR WITHIN THE DEVICE EXHIBITED AN OUT-OF-LIMIT (HIGHER) LEAKAGE CURRENT AT TEST CHAMBER TEMPERATURE. A BATTERY LONGEVITY PREDICTION PERFORMED BY THE MANUFACTURER, WHICH RESULTED IN -1.07 YEARS UNTIL ERI=YES, CONFIRMED THAT THE END OF SERVICE CONDITION WAS AN EXPECTED EVENT. THOUGH THE TRANSISTOR COULD HAVE CONTRIBUTED TO THE END OF THE SERVICE OF THE DEVICE, THIS SLIGHT LEAKAGE WOULD HAVE ONLY HAD A MINIMAL EFFECT ON THE EOS CONDITION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PULSE GEN MODEL 101 LYJ CYBERONICS, INC. 101 3673

Patients

Seq Age Sex Outcome Treatment
1 47 YR