FDA Adverse Event
Malfunction
Summary report: N
PULSE GEN MODEL 101
MDR report key: 1181372
·
Received September 23, 2008
Report
- Report Number
- 1644487-2008-02296
- Event Type
- Malfunction
- Date Received
- September 23, 2008
- Date of Event
- July 26, 2001
- Report Date
- September 4, 2008
- Manufacturer
- CYBERONICS, INC.
- Product Code
- LYJ
- PMA / PMN Number
- P970003
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
Description of Event or Problem · 1
THE REPORTER INDICATED THAT IT WAS DURING THE ANALYSIS OF A VNS PATIENT'S GENERATOR, WHO UNDERWENT REVISION SURGERY DUE TO THE EXPECTED END OF SERVICE OF THE DEVICE, IT WAS REVEALED THAT A TRANSISTOR WITHIN THE DEVICE EXHIBITED AN OUT-OF-LIMIT (HIGHER) LEAKAGE CURRENT AT TEST CHAMBER TEMPERATURE. A BATTERY LONGEVITY PREDICTION PERFORMED BY THE MANUFACTURER, WHICH RESULTED IN -1.07 YEARS UNTIL ERI=YES, CONFIRMED THAT THE END OF SERVICE CONDITION WAS AN EXPECTED EVENT. THOUGH THE TRANSISTOR COULD HAVE CONTRIBUTED TO THE END OF THE SERVICE OF THE DEVICE, THIS SLIGHT LEAKAGE WOULD HAVE ONLY HAD A MINIMAL EFFECT ON THE EOS CONDITION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | PULSE GEN MODEL 101 | LYJ | CYBERONICS, INC. | 101 | 3673 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 47 YR |