FDA Adverse Event Malfunction Summary report: N

KIT,OPEN HEART CON (HCOB )933

MDR report key: 11812099 · Received May 12, 2021

Report

Report Number
1423537-2021-00635
Event Type
Malfunction
Date Received
May 12, 2021
Date of Event
April 26, 2021
Report Date
May 12, 2021
Manufacturer
MEX03 MEXICO-JUAREZ PRESOURCE
Product Code
OEZ
UDI-DI
10888439839852
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
501

Narratives

Additional Manufacturer Narrative · 1

BASED ON SUPPLIER INVESTIGATION, DEVICE HISTORY RECORD (DHR) REVIEW DID NOT INDICATE ANY EXCEPTIONS THAT COULD LEAD TO THE REPORTED INCIDENT. THE AVERAGE LINTING DATA IS 0.157G/10PCS. NO SAMPLE WAS AVAILABLE AT THE TIME OF INVESTIGATION. ACCORDING TO SUPPLIER, OR TOWEL IS MADE OF COTTON, SO COTTON FIBER IS BORN. SUPPLIER CONTINUOUSLY WORKING WITH CARDINAL HEALTH TO BETTER CONTROL THE LINTING AND HAVE IMPLEMENTED SEVERAL MEASURES TO IMPROVE IT: SUCTION MACHINES HAVE BEEN INSTALLED IN GREY CLOTH ROLLING PROCESS, DYEING PROCESS AND CUTTING PROCESS. THE SUCTION PROCESS WAS ADDED BEFORE PRODUCT'S FINAL FOLDING, AND WORKERS DO IT ACCORDING TO STANDARD OPERATION PROCEDURE REQUIREMENT. LINTING TEST METHOD AND ACCEPTABLE CRITERIA WAS STIPULATED TO SEE THE SUCTION RESULTS. (=(B)(4)/(B)(4)PIECES). IN THE FOLDING PROCESS, SUPPLIER USED ONE CLOTH PAD UNDER (B)(4)PIECES SEMI-FINISHED PRODUCTS TO AVOID LINTING STUCK ONTO THE PRODUCTS DURING PRODUCT'S TRANSFER. THE ROOT CAUSE COULD NOT BE DETERMINED. THE COMPLAINT INFORMATION WAS SHARED WITH THE RELEVANT SECTORS FOR THEIR AWARENESS. THERE IS NO ACTION TAKEN AT THIS TIME, BUT THE SUPPLIER WILL CONTINUE TO MONITOR THE TREND OF THIS TYPE OF INCIDENT.

Description of Event or Problem · 1

CUSTOMER REPORTED LINT ON THE BLUE TOWELS PWTB04-STM FROM OPEN HEART KIT, PVCGOHCOB. LINT WAS FOUND PRIOR TO PROCEDURE WITH NO DELAY OR AFFECT TO PATIENT. NO CLINICAL DATA OR PATIENT DEMOGRAPHICS WERE PROVIDED AFTER NUMEROUS ATTEMPTS TO OBTAIN.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
710429 KIT,OPEN HEART CON (HCOB )933 CARDIOVASCULAR PROCEDURE KIT OEZ MEX03 MEXICO-JUAREZ PRESOURCE PVCGOHCOB 554772 10888439839852

Patients

Seq Age Sex Outcome Treatment
1 Other