COLLEAGUE VOLUMETRIC INFUSION PUMP
Report
- Report Number
- 6000001-2007-01889
- Event Type
- Malfunction
- Date Received
- October 2, 2008
- Date of Event
- December 1, 2006
- Report Date
- December 19, 2006
- Manufacturer
- BAXTER HEALTHCARE PTE. LTD.
- Product Code
- FRN
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- OTHER
Narratives
THIS REPORT IS BEING RESUBMITTED IN ACCORDANCE WITH INSTRUCTIONS FROM FDA TO ADDRESS A MANUFACTURER REPORT SEQUENCE NUMBER ISSUE THAT OCCURRED WHEN THIS MDR OR SUPPLEMENT WAS ORIGINALLY SUBMITTED TO THE FDA ON JAN 17 2007. EVALUATION SUMMARY: THE REPORTED CONDITION OF FAILURE CODE 538:320:831:0000 WAS CONFIRMED IN THE PUMP'S EVENT HISTORY FILE DURING PRODUCT EVALUATION. THIS FAILURE CODE WAS CAUSED BY A FAULTY USER INTERFACE MODULE PRINTED CIRCUIT BOARD. THE USER INTERFACE MODULE PRINTED CIRCUIT BOARD WAS REPLACED. REVIEW OF THE COMPLAINT HISTORY REVEALS SIMILAR REPORTS HAVE BEEN RECEIVED FOR THIS PRODUCT FOR THE REPORTED ISSUE. THIS ISSUE IS CURRENTLY BEING INVESTIGATED UNDER CAPA.
THE FACILITY REPORTED A PUMP WITH FAILURE CODE 538:320:831:0000. THIS FAILURE CODE OCCURRED DURING BIO-MED TESTING. THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION NECESSARY ACCORDING TO THE HOSPITAL REPRESENTATIVE. NO ADDITIONAL INFORMATION IS AVAILABLE.