STARCLOSE VASCULAR CLOSURE SYSTEM
Report
- Report Number
- 2953144-2008-01591
- Event Type
- Injury
- Date Received
- September 30, 2008
- Date of Event
- September 12, 2008
- Report Date
- September 12, 2008
- Manufacturer
- ABBOTT VASCULAR REDWOOD CITY
- Product Code
- MGB
- PMA / PMN Number
- P050007
- Removal / Correction Number
- NA
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- PHYSICIAN
Narratives
THE DEVICE IS EXPECTED TO BE RETURNED FOR EVALUATION. IT HAS NOT YET BEEN RECEIVED.
DEVICE MALFUNCTION: CLIP MISLOCATION. TIME OF DEVICE MALFUNCTION: DURING VESSEL CLOSURE. SYMPTOMS/AE: FAILURE TO ACHIEVE HEMOSTASIS. IT WAS REPORTED THAT A FELLOW, WHO IS NOT TRAINED IN THE USE OF THE STARCLOSE DEVICE, BUT WAS BEING MONITORED BY A PHYSICIAN TRAINED IN THE USE OF THE STARCLOSE DEVICE ATTEMPTED ARTERIOTOMY CLOSURE OF THE RIGHT COMMON FEMORAL ARTERY AFTER A DIAGNOSTIC PROCEDURE. REPORTEDLY, THE TRIGGER WAS PRESSED TWICE BEFORE THE CLIP WOULD DEPLOY RESULTING IN THE CLIP DEPLOYING IN THE DISTAL END OF THE DEVICE. THE DEVICE WAS REMOVED AND HEMOSTASIS WAS ACHIEVED USING MANUAL COMPRESSION. THERE WERE NO REPORTED ADVERSE PATIENT EFFECTS. THOUGH REQUESTED, NO ADDITIONAL INFORMATION WAS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | STARCLOSE VASCULAR CLOSURE SYSTEM | MGB | ABBOTT VASCULAR REDWOOD CITY | NA | UNK |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 60 YR | Required Intervention |