FDA Adverse Event Malfunction Summary report: N

LIBRE SENSOR FREESTYLE

MDR report key: 11811000 · Received May 12, 2021

Report

Report Number
2954323-2021-68974
Event Type
Malfunction
Date Received
May 12, 2021
Date of Event
May 6, 2021
Report Date
July 12, 2021
Manufacturer
ABBOTT DIABETES CARE INC
Product Code
PZE
PMA / PMN Number
P160030
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

SENSOR (B)(6) HAS BEEN RETURNED AND INVESTIGATED. THE SENSOR PLUG IS FULLY SEATED AND NO ISSUES WERE OBSERVED. EXTRACTED DATA FROM THE RETURNED SENSOR USING APPROVED SOFTWARE. THE SENSOR WAS FOUND TO BE IN SENSOR STATE 5 (INDICATING NORMAL TERMINATION). INSPECTED THE PLUG ASSEMBLY, NO ISSUES WERE OBSERVED. THE CURRENT WAS APPLIED TO THE SENSOR TO PERFORM LINEARITY TESTING WHILE IN THE TEST FIXTURE. ALL RESULTS WERE WITHIN SPECIFICATION. NO MALFUNCTION OR PRODUCT DEFICIENCY WAS IDENTIFIED. ALL PERTINENT INFORMATION AVAILABLE TO ABBOTT DIABETES CARE HAS BEEN SUBMITTED.

Description of Event or Problem · 0

A CUSTOMER REPORTED RECEIVING ERRONEOUS GLUCOSE RESULTS FROM AN ABBOTT DIABETES CARE DEVICE. THE RESULTS WHEN PLOTTED ON A PARKES ERROR GRID FELL INTO EITHER THE C, D, OR E ZONE. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY, OR MISTREATMENT ASSOCIATED WITH THIS EVENT.

Additional Manufacturer Narrative · 1

AT THIS TIME PRODUCT HAS NOT YET BEEN RETURNED. AN EXTENDED INVESTIGATION HAS BEEN PERFORMED FOR THE REPORTED COMPLAINT AND THERE WAS NO INDICATION THAT THE PRODUCT DID NOT MEET SPECIFICATION. DHRS FOR THE FS LIBRE SENSOR AND FS LIBRE SENSOR KIT WERE REVIEWED AND THE DHRS SHOWED THE FS LIBRE SENSOR AND SENSOR KIT PASSED ALL TESTS PRIOR TO RELEASE. IF THE PRODUCT IS RETURNED, A PHYSICAL INVESTIGATION WILL BE PERFORMED AND A FOLLOW-UP REPORT SUBMITTED. ALL PERTINENT INFORMATION AVAILABLE TO ABBOTT DIABETES CARE HAS BEEN SUBMITTED.

Description of Event or Problem · 1

A CUSTOMER REPORTED RECEIVING ERRONEOUS GLUCOSE RESULTS FROM AN ABBOTT DIABETES CARE DEVICE. THE RESULTS WHEN PLOTTED ON A PARKES ERROR GRID FELL INTO EITHER THE C, D, OR E ZONE. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY, OR MISTREATMENT ASSOCIATED WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
711300 LIBRE SENSOR FREESTYLE FLASH GLUCOSE MONITORING SYSTEM PZE ABBOTT DIABETES CARE INC 71938-01

Patients

Seq Age Sex Outcome Treatment
1