FDA Adverse Event
Malfunction
Summary report: N
PANORAMA
MDR report key: 1181053
·
Received September 29, 2008
Report
- Report Number
- 2221819-2008-00043
- Event Type
- Malfunction
- Date Received
- September 29, 2008
- Date of Event
- July 25, 2008
- Report Date
- July 25, 2008
- Manufacturer
- MINDRAY DS USA, INC.
- Product Code
- DRT
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE COMPANY REPRESENTATIVE DETERMINED THAT THE SYSTEM SERVER HAD CRASHED. THIS OCCURRED DURING REINSTALL OF EMERGENCY EQUIPMENT; HOWEVER, IT DOES NOT APPEAR THAT THE ONGOING INSTALLATION ACTIVITY WAS RELATED TO THE SYSTEM GOING DOWN. THE COMPANY REPRESENTATIVE CLEARED THE SERVER ERROR LOGS AND RESET THE SYSTEM. THE SYSTEM WAS TESTED TO FACTORY SPECIFICATION. IT FUNCTIONED NORMALLY AND WAS RETURNED TO USE.
Description of Event or Problem · 1
THE CUSTOMER REPORTED THAT WHILE THE PANORAMA CENTRAL STATION WAS IN USE MONITORING PTS ALONG WITH TELEPACKS AND PASSPORT 2 MONITORS AT THE BEDSIDES, THE SYSTEM WENT DOWN. NO PT INJURY WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | PANORAMA | PATIENT MONITORING NETWORK | DRT | MINDRAY DS USA, INC. | PANORAMA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK |