FDA Adverse Event Malfunction Summary report: N

BD ULTRA FINE PEN NEEDLES

MDR report key: 11810515 · Received May 12, 2021

Report

Report Number
9616656-2021-00511
Event Type
Malfunction
Date Received
May 12, 2021
Date of Event
April 13, 2021
Report Date
June 7, 2021
Manufacturer
BECTON DICKINSON AND CO.
Product Code
FMI
UDI-DI
00382903205509
PMA / PMN Number
K162516
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

H6: INVESTIGATION SUMMARY: CUSTOMER RETURNED (2) OPEN 4MM, 32G PEN NEEDLES (1 WITHOUT THE TEAR DROP LABEL) FROM LOT # 0203824. CUSTOMER STATES THAT THE NEEDLE CLOGGED DURING INJECTION. BOTH RETURNED PEN NEEDLES WERE EXAMINED AND ONE EXHIBITED A BENT NON PATIENT END OF THE CANNULA. THE REMAINING SAMPLE EXHIBITED A BROKEN NON PATIENT END OF THE CANNULA. A LOT HISTORY REVIEW WAS CARRIED OUT AND NO RELATED NON CONFORMANCES WERE RAISED IN ASSOCIATION WITH THIS PACKAGED LOT CONCLUDING ALL INSPECTIONS WERE PERFORMED AS PER THE APPLICABLE OPERATIONS AND MET QC SPECIFICATIONS. BD WAS ABLE TO DUPLICATE OR CONFIRM THE CUSTOMER¿S INDICATED FAILURE. ROOT CAUSE IS USER RELATED. THE BENT AND BROKEN NON PATIENT END OF THE CANNULA OCCURRED DURING USE OF THE PRODUCT BY THE CUSTOMER.

Description of Event or Problem · 0

IT WAS REPORTED THAT 1 BD ULTRA FINE¿ PEN NEEDLE 4MM X 32G WAS UNABLE TO DELIVER INSULIN/MEDICATION. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER : MATERIAL NO. 320550; BATCH NO. 0203824. IT WAS REPORTED THAT THE NEEDLE CLOGGED DURING INJECTION AND THAT THERE IS NO INSULIN FLOW. DATE OF EVENT: UNKNOWN. SAMPLES: AVAILABLE: SENDING MAIL KIT.

Additional Manufacturer Narrative · 1

DATE OF EVENT: UNKNOWN. THE DATE RECEIVED BY MANUFACTURER HAS BEEN USED FOR THIS FIELD. A DEVICE EVALUATION IS PENDING BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 1

IT WAS REPORTED THAT 1 BD ULTRA FINE¿ PEN NEEDLE 4MM X 32G WAS UNABLE TO DELIVER INSULIN/MEDICATION. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER : MATERIAL NO. 320550, BATCH NO. 0203824. IT WAS REPORTED THAT THE NEEDLE CLOGGED DURING INJECTION AND THAT THERE IS NO INSULIN FLOW. DATE OF EVENT: UNKNOWN. SAMPLES: AVAILABLE - SENDING MAIL KIT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
710044 BD ULTRA FINE PEN NEEDLES HYPODERMIC SINGLE LUMEN NEEDLE FMI BECTON DICKINSON AND CO. 320550 0203824 00382903205509

Patients

Seq Age Sex Outcome Treatment
1