FDA Adverse Event Malfunction Summary report: N

INTELECT MOBILE COMBO

MDR report key: 1180923 · Received September 29, 2008

Report

Report Number
1022819-2008-00232
Event Type
Malfunction
Date Received
September 29, 2008
Date of Event
December 4, 2006
Report Date
December 4, 2006
Manufacturer
CHATTANOOGA GROUP
Product Code
IPF
Removal / Correction Number
1022819-11/15/2007-001-C
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IN
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

THIS DEVICE IS FOR EXPORT ONLY.

Description of Event or Problem · 1

THE DEVICE HAS A HIGH OUTPUT ON CHANNEL 2 WHEN INTENSITY IS RAISED TO 0.1MA. THE DEVICE SURGED/SHOCKED. THIS MALFUNCTION CAUSES THE PRINTED CIRCUIT BOARD TO CHANGE THE STIMULATORY PT OUTPUT FROM THE INTENDED OUTPUT TO AN UNINTENDED HIGHER VOLTAGE OUTPUT. WHEN SUCH A MALFUNCTION OCCURS, IT RESULTS IN A SHOCK SENSATION AND POSSIBLE BURN TO THE AREA BENEATH THE ELECTRODE PADS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 INTELECT MOBILE COMBO IPF, IMG, GZJ, GZI IPF CHATTANOOGA GROUP 2778

Patients

Seq Age Sex Outcome Treatment
1