FDA Adverse Event
Malfunction
Summary report: N
INTELECT ADVANCED 2CH COLOR COMBO W/EMG
MDR report key: 1180919
·
Received September 29, 2008
Report
- Report Number
- 1022819-2008-00235
- Event Type
- Malfunction
- Date Received
- September 29, 2008
- Date of Event
- May 15, 2007
- Report Date
- May 15, 2007
- Manufacturer
- CHATTANOOGA GROUP
- Product Code
- IPF
- Removal / Correction Number
- 1022819-11/15/2007-001-C
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SF
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THIS DEVICE IS FOR EXPORT ONLY.
Description of Event or Problem · 1
PT SHOCKED DURING TREATMENT ON CHANNEL 1. THIS MALFUNCTION CAUSES THE PRINTED CIRCUIT BOARD TO CHANGE THE STIMULATORY PT OUTPUT FROM THE INTENDED OUTPUT TO AN UNINTENDED HIGHER VOLTAGE OUTPUT. WHEN SUCH A MALFUNCTION OCCURS, IT RESULTS IN A SHOCK SENSATION AND POSSIBLE BURN TO THE AREA BENEATH THE ELECTRODE PADS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | INTELECT ADVANCED 2CH COLOR COMBO W/EMG | IPF, IMG, GZJ, HCC, GZI | IPF | CHATTANOOGA GROUP | 2762CC |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |