FDA Adverse Event Malfunction Summary report: N

INTELECT MOBILE 2 CH STIM

MDR report key: 1180912 · Received September 29, 2008

Report

Report Number
1022819-2008-00224
Event Type
Malfunction
Date Received
September 29, 2008
Date of Event
April 24, 2007
Report Date
April 24, 2007
Manufacturer
CHATTANOOGA GROUP
Product Code
IPF
Removal / Correction Number
1022819-11/15/2007-001-C
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TU
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THIS DEVICE IS FOR EXPORT ONLY.

Description of Event or Problem · 1

DEVICE SHOCKED PT DURING ELECTROTHERAPY TREATMENT. THIS MALFUNCTION CAUSES THE PRINTED CIRCUIT BOARD TO CHANGE THE STIMULATORY PT OUTPUT FROM THE INTENDED OUTPUT TO AN UNINTENDED HIGHER VOLTAGE OUTPUT. WHEN SUCH A MALFUNCTION OCCURS, IT RESULTS IN A SHOCK SENSATION AND POSSIBLE BURN.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 INTELECT MOBILE 2 CH STIM IPF, GZJ, GZI IPF CHATTANOOGA GROUP 2777

Patients

Seq Age Sex Outcome Treatment
1