FDA Adverse Event Malfunction Summary report: N

INTELECT ADVANCED STIM MONOCHROME

MDR report key: 1180906 · Received September 29, 2008

Report

Report Number
1022819-2008-00223
Event Type
Malfunction
Date Received
September 29, 2008
Date of Event
April 24, 2007
Report Date
April 24, 2007
Manufacturer
CHATTANOOGA GROUP
Product Code
IPF
Removal / Correction Number
1022819-11/15/2007-001-C
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TU
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THIS DEVICE IS FOR EXPORT ONLY.

Description of Event or Problem · 1

DEVICE SHOCKED PT. THIS MALFUNCTION CAUSES THE PRINTED CIRCUIT BOARD TO CHANGE THE STIMULATORY PT OUTPUT FROM THE INTENDED OUTPUT TO AN UNINTENDED HIGHER VOLTAGE OUTPUT. WHEN SUCH A MALFUNCTION OCCURS, IT RESULTS IN A SHOCK SENSATION AND POSSIBLE BURN TO THE AREA BENEATH THE ELECTRODE PADS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 INTELECT ADVANCED STIM MONOCHROME IPF, GZJ, GZI IPF CHATTANOOGA GROUP 2773MS

Patients

Seq Age Sex Outcome Treatment
1