FDA Adverse Event Malfunction Summary report: N

LIFESHIELD SYMBIQ SET WITH 2 CLAVES

MDR report key: 1180851 · Received September 23, 2008

Report

Report Number
9615050-2008-00280
Event Type
Malfunction
Date Received
September 23, 2008
Date of Event
August 1, 2008
Report Date
August 27, 2008
Manufacturer
HOSPIRA DE COSTA RICA LTD.
Product Code
FRN
PMA / PMN Number
K041550
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
KY, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE CUSTOMER CONTACT INDICATED THE DEVICE WAS DISCARDED.

Description of Event or Problem · 1

THE CUSTOMER CONTACT REPORTED CONCURRENT DELIVERY. THE TUBING SET WAS BEING USED TO DELIVER NORMAL SALINE AT A RATE OF 20ML/HR ON THE PRIMARY LINE VIA A SYMBIQ PUMP. AT AN UNSPECIFIED TIME, A SECONDARY TUBING SET WAS CONNECTED TO THE PRIMARY TUBING SET TO DELIVER AN UNSPECIFIED CONCENTRATION OF KEFZOL AT A RATE OF 120ML/HR. THE PRIMARY SOLUTION CONTAINER WAS LOWERED BELOW THE SECONDARY CONTAINER. AT THIS TIME, THE NURSE REPORTED THE PRIMARY AND SECONDARY SOLUTIONS WERE DELIVERING CONCURRENTLY. THE PRIMARY SOLUTION CONTAINER WAS LOWERED FURTHER BELOW THE SECONDARY SOLUTION CONTAINER AND THE SECONDARY SOLUTION FLOWED. THERE WERE NO REPORTED ADVERSE PATIENT EFFECTS. NO MEDICAL INTERVENTIONS WERE REQUIRED. THOUGH REQUESTED, NO ADDITIONAL INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 LIFESHIELD SYMBIQ SET WITH 2 CLAVES 80-FRN FRN HOSPIRA DE COSTA RICA LTD. NA UNK

Patients

Seq Age Sex Outcome Treatment
1 ADULT LIFESHIELD SECONDARY SET